| ▶ 유/무선 통신기기 |
| 1999/519/EC |
COUNCIL RECOMMENDATION of 12 July 1999 on the limitation of exposure of the general public to electromagnetic fields (0 ㎐ to 300 ㎓) |
0 ㎐ ~ 40 ㎓ |
소재지 |
| ACMA Radiocommunications Standard 2014 |
Radiocommunications (Electromagnetic Radiation – Human Exposure) Standard |
300 ㎒ ~ 6 ㎓ |
소재지 |
| ANSI/IEEE C95.1:2005 |
IEEE Standard for Safety Levels with Respect to Human Exposure to Radio Frequency Electromagnetic Fields, 3 ㎑ to 300 ㎓ |
3 ㎑ ~ 40 ㎓ |
소재지 |
| ARPANSA Radiation Protection Series S-1 |
Maximum Exposure Levels to Radiofrequency Fields - 3 ㎑ to 300 ㎓ |
300 ㎒ ~ 6 ㎓ |
소재지 |
| ARPANSA Radiation Protection Standard |
Maximum Exposure Levels to Radiofrequency Fields - 3 ㎑ to 300 ㎓ |
300 ㎒ ~ 6 ㎓ |
소재지 |
| AS/NZS 2772.2:2016 |
Principles and methods of measurement and computation—3 ㎑ to 300 ㎓ |
3 ㎑ ~ 40 ㎓ |
소재지 |
| AS/NZS 4268:2017 |
Radio equipment and systems - Short range devices - Limits and methods of measurement |
9 ㎑ ~ 40 ㎓ |
소재지 |
| AS/NZS 4268:2017 Amd 1:2021 |
Radio equipment and systems - Short range devices - Limits and methods of measurement |
9 ㎑ ~ 40 ㎓ |
소재지 |
| EN 50360:2001+A1:2012 |
Product standard to demonstrate the compliance of mobile phones with the basic restrictions related to human exposure to electromagnetic fields (300 ㎒ ~ 3 ㎓) |
300 ㎒ ~ 3 ㎓ |
소재지 |
| EN 50360:2017 |
Product standard to demonstrate the compliance of wireless communication devices, with the basic restrictions and exposure limit values related to human exposure to electromagnetic fields in the frequency range from 300 ㎒ to 6 ㎓: devices used next to the ear |
300 ㎒ ~ 6 ㎓ |
소재지 |
| EN 50566:2013+AC:2014 |
Product standard to demonstrate compliance of radio frequency fields from handheld and body-mounted wireless communication devices used by the general public (30 ㎒ ~ 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| EN 50566:2017 |
Product standard to demonstrate the compliance of wireless communication devices with the basic restrictions and exposure limit values related to human exposure to electromagnetic fields in the frequency range from 30 ㎒ to 6 ㎓: hand-held and body mounted devices in close proximity to the human body |
300 ㎒ ~ 6 ㎓ |
소재지 |
| EN 50663:2017 |
Generic standard for assessment of low power electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (10 ㎒ – 300 ㎓) |
10 ㎒ ~ 40 ㎓ |
소재지 |
| EN 62209-1:2006 |
Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices. Human models, instrumentation, and procedures - Part 1: Procedure to determine the specific absorption rate (SAR) for hand-held devices used in close proximity to the ear (frequency range of 300 ㎒ to 3 ㎓) |
300 ㎒ ~ 2.6 ㎓ |
소재지 |
| EN 62209-1:2016 |
Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Part 1: Devices used next to the ear (Frequency range of 300 ㎒ to 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| EN 62209-2:2010 |
Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Human models, instrumentation, and procedures - Part 2: Procedure to determine the specific absorption rate (SAR) for wireless communication devices used in close proximity to the human body(frequency range of 30 ㎒ to 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| EN 62209-2:2010+A1:2019 |
Human exposure to radio frequency fields from handheld and bodymounted wireless communication devices — Human models, instrumentation, and procedures Part 2: Procedure to determine the specific absorption rate (SAR) for wireless communication devices used in close proximity to the human body (frequency range of 30 ㎒ to 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| EN 62311:2008 |
Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0 ㎐ ~ 300 ㎓) |
SAR: 300 ㎒ ~ 6 ㎓ MPE: 0 ㎐ ~ 40 ㎓ |
소재지 |
| EN 62479:2010 |
Assessment of the compliance of low power electronic and electrical equipment with the basic restrictions related to human exposure to electromagnetic fields (10 ㎒ to 300 ㎓) |
10 ㎒ ~ 40 ㎓ |
소재지 |
| EN IEC 62311:2020 |
Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields(0 ㎐ to 300 ㎓) |
SAR: 300 ㎒ ~ 6 ㎓ MPE: 0 ㎐ ~ 40 ㎓ |
소재지 |
| ETSI EN 300 330:2017 |
Short Range Devices (SRD); Radio equipment in the frequency range 9 ㎑ to 25 ㎒ and inductive loop systems in the frequency range 9 ㎑ to 30 ㎒; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
9 ㎑ ~ 18 ㎓ |
소재지 |
| ETSI EN 301 091-1:2017 |
Short Range Devices; Transport and Traffic Telematics (TTT); Radar equipment operating in the 76 ㎓ to 77 ㎓ range; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU; Part 1: Ground based vehicular radar |
30 ㎒ ~ 40 ㎓ |
소재지 |
| ETSI EN 302 264:2017 |
Short Range Devices; Transport and Traffic Telematics (TTT); Short Range Radar equipment operating in the 77 ㎓ to 81 ㎓ band; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
30 ㎒ ~ 40 ㎓ |
소재지 |
| ETSI EN 303 417:2017 |
Wireless power transmission systems, using technologies other than radio frequency beam in the 19 - 21 ㎑, 59 - 61 ㎑, 79 - 90 ㎑, 100 - 300 ㎑, 6 765 - 6 795 ㎑ ranges; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
9 ㎑ ~ 18 ㎓ |
소재지 |
| FCC Part 2.1091:2019 |
Radiofrequency radiation exposure evaluation; mobile devices |
3 ㎑ ~ 40 ㎓ |
소재지 |
| FCC Part 2.1093:2019 |
Radiofrequency radiation exposure evaluation; portable devices |
100 ㎒ ~ 6 ㎓ |
소재지 |
| IEC 62209-1:2005 |
Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices. Human models, instrumentation, and procedures - Part 1: Procedure to determine the specific absorption rate (SAR) for hand-held devices used in close proximity to the ear(frequency range of 300 ㎒ to 3 ㎓) |
300 ㎒ ~ 3 ㎓ |
소재지 |
| IEC 62209-1:2016 |
Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Part 1: Devices used next to the ear (Frequency range of 300 ㎒ to 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| IEC 62209-2:2010 |
Human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices - Human models, instrumentation, and procedures - Part 2: Procedure to determine the specific absorption rate (SAR) for wireless communication devices used in close proximity to the human body(frequency range of 30 ㎒ to 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| IEC 62209-2:2010+AMD1:2019 |
Human exposure to radio frequency fields from handheld and bodymounted wireless communication devices — Human models, instrumentation, and procedures Part 2: Procedure to determine the specific absorption rate (SAR) for wireless communication devices used in close proximity to the human body (frequency range of 30 ㎒ to 6 ㎓) |
300 ㎒ ~ 6 ㎓ |
소재지 |
| IEEE Std 1528™:2013 |
IEEE Recommended Practice for Determining the Peak Spatial - Average Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement Techniques |
300 ㎒ ~ 6 ㎓ |
소재지 |
| IEEE Std C95.3™:2002 |
IEEE Recommended Practice for Measurements and Computations of Radio Frequency Electromagnetic Fields With Respect to Human Exposure to Such Fields, 100 ㎑ - 300 ㎓ |
300 ㎒ ~ 2.6 ㎓ |
소재지 |
| RSS-102 Issue 5:2015 |
Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Bands) |
SAR: 300 ㎒ ~ 6 ㎓ MPE: 3 ㎑ ~ 40 ㎓ |
소재지 |
| TEC/GR/SAR/001/01.MAR-09 |
GR on “SAR (Specific absorption rate) Measurement System” |
300 ㎒ ~ 3 ㎓ |
소재지 |
| ▶ 가정용 전기기기 |
| EN 50332-1:2013 |
Sound system equipment: Headphones and earphones associated with portable audio equipment - Maximum sound pressure level measurement methodology and limit considerations - Part 1: General method for "one package equipment" |
< 100 ㏈ |
소재지-1 |
| EN 50332-2:2013 |
Sound system equipment: Headphones and earphones associated with portable audio equipment - Maximum sound pressure level measurement methodology and limit considerations - Part 2: Matching of sets with headphones if either or both are offered separately |
< 150 ㎷ |
소재지-1 |
| EN 62368-1:2014+A11:2017 |
Audio/video, information and communication technology equipment - Part 1: Safety requirements <제외항목> 4.4.4.6 Glass impact test 5.5.2 Capacitors and RC 5.5.3 Transformers 5.5.4 Optocouplers 5.5.5 Relays 5.5.6 Resistors 5.5.7 SPD 5.5.8 Insulation between the mains and an external circuit consisting of a coaxial cable 8.5.5 High pressure lamps 10.3 Safeguards against laser radiation 10.5 Safeguards against X-radiation Annex C UV radiation Annex D Test generators Annex G.7 Mains supply cords Annex G.9 Integrated circuit (IC) current limiters Annex G.13.6 Tests on coated printed boards Annex G.15 Pressurized liquid filled components Annex G.16 IC including capacitor discharge function (ICX) Annex H Criteria for telephone ringing signals Annex J Insulated winding wires for use without interleaved insulation Annex K Safety interlocks Annex M.7 Risk of explosion from lead acid and NiCd batteries Annex M.8 Protection against internal ignition from external spark sources of batteries with aqueous electrolyte Annex R Limited shortcircuit test Annex S.3 Flammability test for bottom of a fire enclosure Annex T.10 Glass fragmentation test Annex U Mechanical strength of CRTs and protection against the effects of implosion |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| EN IEC 62368-1:2020+A11:2020 |
Audio/video, information and communication technology equipment - Part 1: Safety requirements <제외항목> 5.5.2 Capacitors and RC 5.5.3 Transformers 5.5.4 Optocouplers 5.5.5 Relays 5.5.6 Resistors 5.5.7 SPD 5.5.8 Insulation between the mains and an external circuit consisting of a coaxial cable 8.5.5 High pressure lamps 10.3 Safeguards against laser radiation 10.5 Safeguards against X-radiation 10.6.3 Requirements for dose-based systems Annex C UV radiation Annex D Test generators Annex G.5.3.4 Transformers using fully insulated winding wire (FIW) Annex G.7 Mains supply cords Annex G.9 Integrated circuit (IC) current limiters Annex G.13.6 Tests on coated printed boards Annex G.15 Pressurized liquid filled components Annex G.16 IC including capacitor discharge function (ICX) Annex H Criteria for telephone ringing signals Annex J Insulated winding wires for use without interleaved insulation Annex K Safety interlocks Annex M.7 Risk of explosion from lead acid and NiCd batteries Annex M.8 Protection against internal ignition from external spark sources of batteries with aqueous electrolyte Annex R Limited shortcircuit test Annex S.3 Flammability test for bottom of a fire enclosure Annex T.10 Glass fragmentation test Annex U Mechanical strength of CRTs and protection against the effects of implosion Annex Y Construction requirements for outdoor enclosures |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| EN IEC 62368-1:2024+A11:2024 |
Audio/video, information and communication technology equipment - Part 1: Safety requirements <제외항목> 5.5.2 Capacitors and RC 5.5.3 Transformers 5.5.4 Optocouplers 5.5.5 Relays 5.5.6 Resistors 5.5.7 Surge suppressors 5.5.8 Insulation between the mains and an external circuit consisting of a coaxial cable 8.5.5 High pressure lamps 10.3 Safeguards against laser radiation 10.5 Safeguards against X-radiation 10.6.3 Requirements for dose-based systems Annex C UV radiation Annex D Test generators Annex G.5.3.4 Transformers using fully insulated winding wire (FIW) Annex G.7 Mains power supply cords and interconnection cables Annex G.9 Integrated circuit (IC) current limiters Annex G.13.6 Tests on coated printed boards Annex G.15 Pressurized liquid filled components or LFC assemblies Annex G.16 IC including capacitor discharge function (ICX) Annex H Criteria for telephone ringing signals Annex J Insulated winding wires for use without interleaved insulation Annex K Safety interlocks Annex M.7 Risk of explosion from lead acid and NiCd batteries Annex M.8 Protection against internal ignition from external spark sources of rechargeable batteries with aqueous electrolyte Annex R Limited shortcircuit test Annex S.3 Flammability test for bottom of a fire enclosure Annex T.10 Glass fragmentation test Annex U Mechanical strength of CRTs and protection against the effects of implosion Annex Y Construction requirements for outdoor enclosures |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 Nm 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| EN IEC 62368-3:2020 |
Audio/video, information and communication technology equipment – Part 3: Safety aspects for DC power transfer through communication cables and ports |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A |
소재지-1 |
| IEC 60065:2014 |
Audio, video and similar electronic apparatus - Safety requirements <제외항목> 6.2 Laser radiation 8.18 Endurance test for wound components with insulated winding wires without additional interleaved insulation 10.1 Surge test 12.1.2 Vibration 14.1 Resistors 14.2 Capacitors and RC-units 14.3 Inductors and windings 14.4 High voltage components and assemblies 14.5 Protective devices 14.6 Switches 14.7 Safety interlock 14.9 Motors 14.12 Surge suppression varistors 15.1 Plugs and sockets 16.1 MAINS supply flexible cords 16.3 Flexible cords 18. Mechanical strength of picture tubes and protection against the effects of implosion 19.6 Mechanical strength of glass Annex A : Additional requirements for apparatus with protection against splashing water Annex H : Insulated winding wires for use without Interleaved insulation |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 60950-1:2005+A1:2009+A2:2013 |
Information technology equipment safety <제외항목> 2.8 Safety interlocks 2.10.5.4 Semiconductor devices 2.10.5.8 Non-separable thin sheet material 2.10.5.12 Wire in wound components 2.10.6.2 Coated printed boards 3.2.5.1 AC Power supply cord 4.2.8 Cathode ray tubes 4.3.12 Flammable liquids 4.3.13 Radiation 4.6.5 Adhesives for constructional purposes 5.3.8 Unattended equipment 7.2 Protection of cable distribution system service persons, and users of other equipment connected to the system, from hazardous voltages in the equipment 7.3 Protection of equipment users from overvoltages on the cable distribution system Annex A Tests for resistance to heat and fire Annex K Thermal controls Annex U Insulated winding wires for use without interleaved insulation Annex Y Ultraviolet light conditioning test Annex AA Mandrel Test |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 60950-21:2002 |
Information technology equipment - Safety - Part 21: Remote power feeding |
인가 전압: 300 V a.c.측정 전압: 600 V a.c.인가 전류: 30 A 측정 전류: 30 A 인가 주파수: (50/60) ㎐ 측정 주파수: 10 ㎒ 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 커패시턴스: 10 F 측정 시간: 1/100 s |
소재지-1 |
| IEC 62368-1:2014 |
Audio/video, information and communication technology equipment - Part 1: Safety requirements <제외항목> 4.4.4.6 Glass impact test 5.5.2 Capacitors and RC 5.5.3 Transformers 5.5.4 Optocouplers 5.5.5 Relays 5.5.6 Resistors 5.5.7 SPD 5.5.8 Insulation between the mains and an external circuit consisting of a coaxial cable 8.5.5 High pressure lamps 10.3 Safeguards against laser radiation 10.5 Safeguards against X-radiation Annex C UV radiation Annex D Test generators Annex G.7 Mains supply cords Annex G.9 Integrated circuit (IC) current limiters Annex G.13.6 Tests on coated printed boards Annex G.15 Pressurized liquid filled components Annex G.16 IC including capacitor discharge function (ICX) Annex H Criteria for telephone ringing signals Annex J Insulated winding wires for use without interleaved insulation Annex K Safety interlocks Annex M.7 Risk of explosion from lead acid and NiCd batteries Annex M.8 Protection against internal ignition from external spark sources of batteries with aqueous electrolyte Annex R Limited shortcircuit test Annex S.3 Flammability test for bottom of a fire enclosure Annex T.10 Glass fragmentation test Annex U Mechanical strength of CRTs and protection against the effects of implosion |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 62368-1:2018 |
Audio/video, information and communication technology equipment - Part 1: Safety requirements <제외항목> 5.5.2 Capacitors and RC 5.5.3 Transformers 5.5.4 Optocouplers 5.5.5 Relays 5.5.6 Resistors 5.5.7 SPD 5.5.8 Insulation between the mains and an external circuit consisting of a coaxial cable 8.5.5 High pressure lamps 10.3 Safeguards against laser radiation 10.5 Safeguards against X-radiation 10.6.3 Requirements for dose-based systems Annex C UV radiation Annex D Test generators Annex G.5.3.4 Transformers using fully insulated winding wire (FIW) Annex G.7 Mains supply cords Annex G.9 Integrated circuit (IC) current limiters Annex G.13.6 Tests on coated printed boards Annex G.15 Pressurized liquid filled components Annex G.16 IC including capacitor discharge function (ICX) Annex H Criteria for telephone ringing signals Annex J Insulated winding wires for use without interleaved insulation Annex K Safety interlocks Annex M.7 Risk of explosion from lead acid and NiCd batteries Annex M.8 Protection against internal ignition from external spark sources of batteries with aqueous electrolyte Annex R Limited shortcircuit test Annex S.3 Flammability test for bottom of a fire enclosure Annex T.10 Glass fragmentation test Annex U Mechanical strength of CRTs and protection against the effects of implosion Annex Y Construction requirements for outdoor enclosures |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 62368-1:2023 |
Audio/video, information and communication technology equipment - Part 1: Safety requirements <제외항목> 5.5.2 Capacitors and RC 5.5.3 Transformers 5.5.4 Optocouplers 5.5.5 Relays 5.5.6 Resistors 5.5.7 Surge suppressors 5.5.8 Insulation between the mains and an external circuit consisting of a coaxial cable 8.5.5 High pressure lamps 10.3 Safeguards against laser radiation 10.5 Safeguards against X-radiation 10.6.3 Requirements for dose-based systems Annex C UV radiation Annex D Test generators Annex G.5.3.4 Transformers using fully insulated winding wire (FIW) Annex G.7 Mains power supply cords and interconnection cables Annex G.9 Integrated circuit (IC) current limiters Annex G.13.6 Tests on coated printed boards Annex G.15 Pressurized liquid filled components or LFC assemblies Annex G.16 IC including capacitor discharge function (ICX) Annex H Criteria for telephone ringing signals Annex J Insulated winding wires for use without interleaved insulation Annex K Safety interlocks Annex M.7 Risk of explosion from lead acid and NiCd batteries Annex M.8 Protection against internal ignition from external spark sources of rechargeable batteries with aqueous electrolyte Annex R Limited shortcircuit test Annex S.3 Flammability test for bottom of a fire enclosure Annex T.10 Glass fragmentation test Annex U Mechanical strength of CRTs and protection against the effects of implosion Annex Y Construction requirements for outdoor enclosures |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 Nm 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 62368-3:2017 |
Audio/video, information and communication technology equipment – Part 3: Safety aspects for DC power transfer through communication cables and ports |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A |
소재지-1 |
| KC 62368-1:2021 |
오디오/비디오 및 정보통신 기술기기 제1부 : 안전 요구 사항 <제외항목> 4.4.3.6 유리 충격 시험 5.5.2 캐패시터 및 RC 장치 5.5.3 변압기 5.5.4 포토 커플러 5.5.5 릴레이 5.5.6 저항기 5.5.7 SPDs 5.5.8 주전원과 동축 케이블로 구성된 외부 회로 사이의 절연 8.5.5 고압 램프 10.3 레이저 방사(선)에 대한 보호수단 10.5 엑스선(X-선)에 대한 보호수단 10.6.3 선량 기반 시스템에 대한 요구 사항 부속서 C 자외선(UV radiation) 부속서 D 시험 발생기 Annex G.5.3.4 FIW(완전 절연 권선)을 사용한 변압기 Annex G.7 주전원 코드 Annex G.9 집적회로(IC) 전류 제한기 Annex G.13.6 코팅 인쇄기판 Annex G.15 가압 액체로 채워진 부품 Annex G.16 캐패시터 방전 기능을 포함한 집적회로(ICX) 부속서 H 전화 벨(링) 신호에 대한 기준 부속서 J 중간절연 없이 사용을 위한 절연권선 전선 부속서 K 안전 인터락 부속서 M.7 납축전지 및 니켈-카드뮴 배터리의 폭발위험 부속서 M.8 수성 전해액 배터리의 외부 스파크 원으로부터 내부 전화에 대한 보호 부속서 R 제한 단락 회로 시험 부속서 S.3 방화용 엔클로우저의 바닥에 대한 난연성 시험 부속서 T.10 유리 파쇄 시험 부속서 U CRTs의 기계적 강도 및 내파 영향에 대한 보호 부속서 Y 옥외용 엔클로우저 구성 요구 사항 |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: (50/60) ㎐ 6. 측정 주파수: 10 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 F 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| ▶ 의료기기 |
| EN 60601-1:2006+A1:2013+AC:2014+A12:2014 |
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance <제외항목> 9.5 Expelled parts Hazard 9.5.2 Cathode ray tubes 9.7.5 Pressure vessels 11.2.2 ME Equipment and ME systems used in conjunction with Oxygen Rich Environments 11.4 ME Equipment and ME Systems intended for use with flammable anaesthetics 11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 11.7 Biocompatibility of ME Equipment and ME Systems 13.2.11 Failures of components in ME EQUIPMENT used in conjunction with OXYGEN RICH ENVIRONMENTS 15.4.3.4 Lithium batteries 15.4.7.3 Entry of liquids |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13.측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| EN 60601-1:2006+A1:2013+A2:2021 |
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance <제외항목> 9.5 Expelled parts Hazard 9.5.2 Cathode ray tubes 9.7.5 Pressure vessels 11.2.2 ME Equipment and ME systems used in conjunction with Oxygen Rich Environments 11.4 ME Equipment and ME Systems intended for use with flammable anaesthetics 11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 11.7 Biocompatibility of ME Equipment and ME Systems 13.2.11 Failures of components in ME EQUIPMENT used in conjunction with OXYGEN RICH ENVIRONMENTS 15.4.3.4 Lithium batteries 15.4.7.3 Entry of liquids |
1.인가 전압: 300 V a.c. 2.측정 전압: 600 V a.c. 3.인가 전류: 30 A 4.측정 전류: 30 A 5.인가 주파수: 50/60 ㎐ 6.측정 주파수: 500 ㎒ 7.측정 전력: 10 ㎾ 8.측정 거리: 1 500 ㎜ 9.인가 온도: - 20 ℃ ~ 150 ℃ 10.측정 온도: 200 ℃ 11.측정 절연저항: ∞ Ω 12.측정 저항: 100 ㏀ 13.측정 커패시턴스: 10 nF 14.인가 에너지: 0.5 J / 5 J 15.측정 토오크: 1.96 N⋅m 16.측정 시간: 1/100 s 17.인가 습도: 95 % R.H. 18.측정 습도: 70 % R.H. 19.난연 인가온도: 960 ℃ 20.측정 각도: 40° 21.측정 무게: 150 ㎏ 22.인가 힘: 5 000 N |
소재지-1 |
| EN 60601-1-10:2008+A1:2015 |
Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers |
- |
소재지-1 |
| EN 60601-1-10:2008+A1:2015+A2:2021 |
Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers |
- |
소재지-1 |
| EN 60601-1-11:2015 |
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment <제외항목> 4.2.3.2 Environmental shock to TRANSIT-OPERABLE ME EQUIPMENT g), I) 8.3.1 Ingress of water or particulate matter in to ME SYSTEMS 8.3.2 Ingress of water or particulate matter into ME SYSTEMS 10.1.2 Requirement for mechanical strength for non-TRANSIT-OPERABLE ME EQUIPMENT 10.1.3 Requirement for mechanical strength for TRANSIT-OPERABLE ME EQUIPMENT |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃인가 힘: 5 000 N 인가 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| EN 60601-1-11:2015+A1:2021 |
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment <제외항목> 4.2.3.2 Environmental shock to TRANSIT-OPERABLE ME EQUIPMENT g), I) 8.3.1 Ingress of water or particulate matter in to ME SYSTEMS 8.3.2 Ingress of water or particulate matter into ME SYSTEMS 10.1.2 Requirement for mechanical strength for non – TRANSIT - OPERABLE ME EQUIPMENT 10.1.3 Requirement for mechanical strength for TRANSIT-OPERABLE ME EQUIPMENT |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 인가 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| EN 60601-1-12:2015 |
Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment <제외항목> 8.1.1 Ingress of water or particulate matter into ME EQUIPMENT 10.1.2 Requirements for mechanical strength for FIXED or PERMANENTLY INSTALLED ME EQUIPMENT intended for use in a road ambulance 10.1.3 Requirements for mechanical strength for TRANSPORTABLE ME EQUIPMENT 10.1.4 Requirements for mechanical strength for ME EQUIPMENT intended for airborne use |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 인가 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| EN 60601-1-12:2015+A1:2020 |
Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment <제외항목> 8.1.1 Ingress of water or particulate matter into ME EQUIPMENT 10.1.2 Requirements for mechanical strength for FIXED or PERMANENTLY INSTALLED ME EQUIPMENT intended for use in a road ambulance 10.1.3 Requirements for mechanical strength for TRANSPORTABLE ME EQUIPMENT 10.1.4 Requirements for mechanical strength for ME EQUIPMENT intended for airborne use |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 인가 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| EN 60601-1-3:2008+A1:2013+AC:2014+A11:2016 |
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-ray equipment |
측정 거리: 1 500 ㎜ 측정 관전압: 50 kVp ~ 150 kVp 측정 조도: 1 500 ㏓ 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN 60601-1-3:2008+A1:2013+A11:2016+A2:2021 |
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-ray equipment |
측정 거리: 1 500 ㎜ 측정 관전압: 50 kVp ~ 150 kVp 측정 조도: 1 500 ㏓ 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN 60601-1-6:2010+A1:2015 |
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
- |
소재지-1 |
| EN 60601-1-6:2010+A1:2015+A2:2021 |
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
- |
소재지-1 |
| EN 60601-1-8:2007+A1:2013+AC:2014+A11:2017 |
Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems |
측정 소음: 94 ㏈ |
소재지-1 |
| EN 60601-1-8:2007+A1:2013+A11:2017+A2:2021 |
Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems |
측정 소음: 94 ㏈ |
소재지-1 |
| EN 60601-1-9:2008+A1:2013 |
Medical electrical equipment - Part 1-9: General requirements for basic safety and essential performance - Collateral standard: General requirements for environmentally conscious design |
- |
소재지-1 |
| EN 60601-1-9:2008+A1:2013+A2:2020 |
Medical electrical equipment - Part 1-9: General requirements for basic safety and essential performance - Collateral standard: General requirements for environmentally conscious design |
- |
소재지-1 |
| EN 60601-2-10:2015+A1:2016 |
Medical electrical equipment - Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 주파수: 500 ㎒ 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 저항: 100 ㏀ |
소재지-1 |
| EN 60601-2-18:2015 |
Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30A 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 용량: 2 000 ㎖ 측정 대기압: 1 030 h㎩ |
소재지-1 |
| EN 60601-2-24:2015 |
Medical electrical equipment - Part 2-24: Particular requirements for the basic safety and essential performance of infusion pumps and controllers <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems |
전자 저울: 0.01 ㎎ |
소재지-1 |
| EN 60601-2-25:2015 |
Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs |
시그널 제너레이터: 20 V peak-to-valley of 10 ㎐ |
소재지-1 |
| EN 60601-2-27:2014 |
Medical electrical equipment - Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 에너지: 600 J |
소재지-1 |
| EN 60601-2-34:2014 |
Medical electrical equipment - Part 2-34: Particular requirements for the safety, including essential performance, of direct blood pressure monitoring equipment <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 201.12.1 Accuracy of controls and instruments |
측정 전압: 5 ㎸ d.c. 측정 소음: 94 ㏈ |
소재지-1 |
| EN 60601-2-36:2015 |
Medical electrical equipment - Part 2-36: Particular requirements for the basic safety and essential performance of equipment for extracorporeally induced lithotripsy <제외항목> 201.11.7 Biocompatibility of ME EQUIPMENT and ME SYSTEMS |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃인가 힘: 5 000 N 인가 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| EN 60601-2-37:2008+A11:2011+A1:2015 |
Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems |
측정 온도: 60 ℃ 측정 주파수: (0.5 ~ 5) ㎒ 측정 출력 세기: 30 W |
소재지-1 |
| EN 60601-2-4:2011 |
Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 201.11.6.7 Sterilization of ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 주파수: 1 000 ㎒ 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 측정 에너지: 600 J 인가 힘: 5 000 N 측정 시간: 1/100 s |
소재지-1 |
| EN 60601-2-4:2011+A1:2019 |
Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 201.11.6.7 Sterilization of ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 주파수: 500 ㎒ 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 측정 에너지: 600 J 인가 힘: 5 000 N 측정 시간: 1/100 s |
소재지-1 |
| EN 60601-2-43:2010+AC:2014+A1:2018 |
Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 40 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN 60601-2-44:2009+A11:2011+A1:2012+A2:2016 |
Medical electrical equipment - Part 2-44: Particular requirements for basic safety and essential performance of X-ray equipment for computed tomography |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 시간: 1/100 s 측정 각도: 40° 측정 무게: 150 ㎏ 측정 관전압: 50 kVp ~ 150 kVp |
소재지-1 |
| EN 60601-2-45:2011+A1:2015 |
Medical electrical equipment - Part 2-45: Particular requirements for basic safety and essential performance of mammographic X-ray equipment and mammographic stereotactic divices <제외항목> 201.9.2.103.3 Biopsy needle positioning accuracy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 시간: 1/100 s 측정 각도: 40° 측정 무게: 150 ㎏ 측정 관전압: 20 kVp ~ 49 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN 60601-2-47:2015 |
Medical electrical equipment - Part 2-47: Particular requirements for the safety the basic safety and essential performance of ambulatory electrocardiographic systems |
주파수 제너레이터: 20 V peak-to-valley of 10 ㎐ |
소재지-1 |
| EN 60601-2-5:2015 |
Medical electrical equipment - Part 2-5: Particular requirements for the basic safety and essential performance of ultrasonic physiotherapy equipment <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems |
측정 온도: 60 ℃ 측정 주파수: (0.5 ~ 5) ㎒ 측정 출력: 30 W |
소재지-1 |
| EN 60601-2-52:2010+A1:2015 |
Medical electrical equipment - Part 2-52: Particular requirements for the basic safety and essential performance of medical beds <제외항목> 201.11.6.5.101 Additional requirements for ingress of water of particulate matter into ME EQUIPMENT and ME SYSTEM |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 측정 커패시턴스: 10 F 측정 에너지: 600 J |
소재지-1 |
| EN 60601-2-54:2009+A1:2015 |
Medical electrical equipment - Part 2-54: Particular requirements for basic safety and essential performance of X-ray equipment for radiography and radioscopy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 40 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN 60601-2-54:2009+A1:2015+A2:2019 |
Medical electrical equipment - Part 2-54: Particular requirements for basic safety and essential performance of X-ray equipment for radiography and radioscopy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 50 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN 60601-2-57:2011 |
Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 커패시턴스: 10 F 측정 에너지: 600 J |
소재지-1 |
| EN 60601-2-62:2015 |
Medical electrical equipment - Part 2-62: Particular requirements for the basic safety and essential performance of high intensity therapeutic ultrasound (HITU) equipment <제외항목> 11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
측정 온도: 60 ℃ 측정 주파수: (0.03 ~ 20) ㎒ 측정 출력 세기: 100 W |
소재지-1 |
| EN 60601-2-63:2015 |
Medical electrical equipment - Part 2-63: Particular requirements for basic safety and essential performance of dental extra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7 m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| EN 60601-2-63:2015+A1:2019 |
Medical electrical equipment - Part 2-63: Particular requirements for basic safety and essential performance of dental extra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7 m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| EN 60601-2-63:2015+A1:2019+A2:2021 |
Medical electrical equipment - Part 2-63: Particular requirements for basic safety and essential performance of dental extra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| EN 60601-2-65:2013 |
Medical electrical equipment - Part 2-65: Particular requirements for basic safety and essential performance of dental intra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7 m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| EN 60601-2-65:2013+A1:2020+A2:2021 |
Medical electrical equipment - Part 2-65: Particular requirements for basic safety and essential performance of dental intra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| EN 62304:2006+A1:2015 |
Medical device software - Software life-cycle processes |
- |
소재지-1 |
| EN 62366:2008+A1:2015 |
Medical devices. Application of usability engineering to medical devices |
- |
소재지-1 |
| EN 62366-1:2015 |
Medical devices - Part 1: Application of usability engineering to medical devices |
- |
소재지-1 |
| EN 62366-1:2015+A1:2020 |
Medical devices - Part 1: Application of usability engineering to medical devices |
- |
소재지-1 |
| EN 80601-2-30:2010+A1:2015 |
Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers <제외항목> 201.15.3.5.101 Shock and vibration for other than transport 211.8.3.1 Ingress of water or particulate matter into ME EQUIPMENT |
최대 측정 압력: 330 ㎜Hg |
소재지-1 |
| EN IEC 60601-2-2:2018 |
Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories <제외항목> 11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 201.15.101 NEUTRAL ELECTRODES |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 용량: 2 000 ㎖ 측정 대기압: 1 030 h㎩ |
소재지-1 |
| EN IEC 60601-2-22:2020 |
Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment |
측정 거리: 1 500 ㎜ 인가 힘: 5 000 N 측정 시간: 1/100 s Laser power: 10 ㎻ ~ 300 W Laser energy: 10 µJ ~ 2 J 측정 파장: (365 ~ 1 100) ㎚ |
소재지-1 |
| EN IEC 60601-2-28:2019 |
Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis |
측정 시간: 1/100 s 측정 저항: 100 ㏀ |
소재지-1 |
| EN IEC 80601-2-30:2019 |
Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers <제외항목> 201.15.3.5.101 Shock and vibration (robustness) 211.8.3.1 Ingress of water or particulate matter into ME EQUIPMENT |
최대 측정 압력: 330 ㎜Hg |
소재지-1 |
| EN IEC 60601-2-43:2023 |
Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 40 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| EN IEC 80601-2-49:2019 |
Medical electrical equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
1. 인가 전압: 300 Va.c. 2. 측정 전압: 600 Va.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7 .측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: (- 20~150) ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| EN IEC 80601-2-60:2020 |
Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 커패시턴스: 10 nF 측정 에너지: 600 J |
소재지-1 |
| IEC 60601-1:2005+A1:2012+A2:2020 |
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance <제외항목> 9.5 Expelled parts Hazard 9.5.2 Cathode ray tubes 9.7.5 Pressure vessels 11.2.2 ME Equipment and Me Systems used in conjunction with Oxygen Rich Environments 11.4 ME Equipment and ME Systems intended for use with flammable anaesthetics 11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 11.7 Biocompatibility of ME Equipment and ME Systems 13.2.11 Failures of components in ME EQUIPMENT used in conjunction with OXYGEN RICH ENVIRONMENTS 15.4.3.4 Lithium Batteries 15.4.7.3 Entry of liquids |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 60601-1:2005+CORR1:2006+CORR2:2007+A1:2012 |
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance <제외항목> 9.5 Expelled parts Hazard 9.5.2 Cathode ray tubes 9.7.5 Pressure vessels 11.2.2 ME Equipment and Me Systems used in conjunction with Oxygen Rich Environments 11.4 ME Equipment and ME Systems intended for use with flammable anaesthetics 11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 11.7 Biocompatibility of ME Equipment and ME Systems 13.2.11 Failures of components in ME EQUIPMENT used in conjunction with OXYGEN RICH ENVIRONMENTS 15.4.3.4 Lithium Batteries 15.4.7.3 Entry of liquids |
1. 인가 전압: 300 V a.c. 2. 측정 전압: 600 V a.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: - 20 ℃ ~ 150 ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 60601-1-10:2007+A1:2013 |
Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers |
- |
소재지-1 |
| IEC 60601-1-10:2007+A1:2013+A2:2020 |
Medical electrical equipment - Part 1-10: General requirements for basic safety and essential performance - Collateral Standard: Requirements for the development of physiologic closed-loop controllers |
- |
소재지-1 |
| IEC 60601-1-11:2015 |
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment <제외항목> 4.2.3.2 Environmental shock to TRANSIT-OPERABLE ME EQUIPMENT g), I) 8.3.1 Ingress of water or particulate matter in to ME SYSTEMS 8.3.2 Ingress of water or particulate matter into ME SYSTEMS 10.1.2 Requirement for mechanical strength for non-TRANSIT-OPERABLE ME EQUIPMENT 10.1.3 Requirement for mechanical strength for TRANSIT-OPERABLE ME EQUIPMENT |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 측정 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| IEC 60601-1-11:2015+A1:2020 |
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment <제외항목> 4.2.3.2 Environmental shock to TRANSIT-OPERABLE ME EQUIPMENT g), I) 8.3.1 Ingress of water or particulate matter in to ME SYSTEMS 8.3.2 Ingress of water or particulate matter into ME SYSTEMS 10.1.2 Requirement for mechanical strength for non-TRANSIT-OPERABLE ME EQUIPMENT 10.1.3 Requirement for mechanical strength for TRANSIT-OPERABLE ME EQUIPMENT |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 측정 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| IEC 60601-1-12:2014 |
Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment <제외항목> 8.1.1 Ingress of water or particulate matter into ME EQUIPMENT 10.1.2 Requirements for mechanical strength for FIXED or PERMANENTLY INSTALLED ME EQUIPMENT intended for use in a road ambulance 10.1.3 Requirements for mechanical strength for TRANSPORTABLE ME EQUIPMENT 10.1.4 Requirements for mechanical strength for ME EQUIPMENT intended for airborne use |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 측정 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| IEC 60601-1-12:2014+A1:2020 |
Medical electrical equipment - Part 1-12: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems intended for use in the emergency medical services environment <제외항목> 8.1.1 Ingress of water or particulate matter into ME EQUIPMENT 10.1.2 Requirements for mechanical strength for FIXED or PERMANENTLY INSTALLED ME EQUIPMENT intended for use in a road ambulance 10.1.3 Requirements for mechanical strength for TRANSPORTABLE ME EQUIPMENT 10.1.4 Requirements for mechanical strength for ME EQUIPMENT intended for airborne use |
인가 전압: 300 V a.c. 인가 전류: 30 A 측정 전류: 30 A 측정 주파수: 500 ㎒ 인가 온도: - 20 ℃ ~ 150 ℃ 인가 힘: 5 000 N 측정 시간: 1/100 s 인가 습도: 95 % R.H. |
소재지-1 |
| IEC 60601-1-3:2008+A1:2013 |
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-ray equipment |
측정 거리: 1 500 ㎜ 측정 관전압: 50 kVp ~ 150 kVp 측정 조도: 1 500 ㏓ 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-1-3:2008+A1:2013+A2:2021 |
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral standard: Radiation protection in diagnostic X-ray equipment |
측정 거리: 1 500 ㎜ 측정 관전압: 50 kVp ~ 150 kVp 측정 조도: 1 500 ㏓ 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-1-6:2010+A1:2013 |
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
- |
소재지-1 |
| IEC 60601-1-6:2010+A1:2013+A2:2020 |
Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability |
- |
소재지-1 |
| IEC 60601-1-8:2006+A1:2012 |
Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems |
측정 소음: 94 ㏈ |
소재지-1 |
| IEC 60601-1-8:2006+A1:2012+A2:2020 |
Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems |
측정 소음: 94 ㏈ |
소재지-1 |
| IEC 60601-1-9:2007+A1:2013 |
Medical electrical equipment - Part 1-9: General requirements for basic safety and essential performance - Collateral standard: General requirements for environmentally conscious design |
- |
소재지-1 |
| IEC 60601-1-9:2007+A1:2013+A2:2020 |
Medical electrical equipment - Part 1-9: General requirements for basic safety and essential performance - Collateral standard: General requirements for environmentally conscious design |
- |
소재지-1 |
| IEC 60601-2-10:2012+A1:2016 |
Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 300 A 측정 주파수: 500 ㎒ 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 저항: 100 ㏀ |
소재지-1 |
| IEC 60601-2-10:2012+A1:2016+A2:2023 |
Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 300 A 측정 주파수: 500 ㎒ 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 저항: 100 ㏀ |
소재지-1 |
| IEC 60601-2-18:2009 |
Medical electrical equipment - Part 2-18: Particular requirements for basic safety and essential performance of endoscopic equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 용량: 2 000 ㎖ 측정 대기압: 1 030 h㎩ |
소재지-1 |
| IEC 60601-2-2:2017 |
Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories <제외항목> 11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 201.15.101 NEUTRAL ELECTRODES |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 용량: 2 000 ㎖ 측정 대기압: 1 030 h㎩ |
소재지-1 |
| IEC 60601-2-2:2017+A1:2023 |
Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories <제외항목> 11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 11.6.7 Sterilization of ME Equipment and ME Systems 201.15.101 NEUTRAL ELECTRODES |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 용량: 2 000 ㎖ 측정 대기압: 1 030 h㎩ |
소재지-1 |
| IEC 60601-2-22:2007+A1:2012 |
Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment |
측정 거리: 1 500 ㎜ 인가 힘: 5 000 N 측정 시간: 1/100 s Laser power: 10 ㎻ ~ 300 W Laser energy: 10 µJ ~ 2 J 측정 파장: (365 ~ 1 100) ㎚ |
소재지-1 |
| IEC 60601-2-22:2019 |
Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment |
측정 거리: 1 500 ㎜ 인가 힘: 5 000 N 측정 시간: 1/100 s Laser power: 10 ㎻ ~ 300 W Laser energy: 10 µJ ~ 2 J 측정 파장: (365 ~ 1 100) ㎚ |
소재지-1 |
| IEC 60601-2-24:2012 |
Medical electrical equipment - Part 2-24: Particular requirements for the basic safety and essential performance of infusion pumps and controllers <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
전자 저울: 0.01 ㎎ |
소재지-1 |
| IEC 60601-2-25:2011 |
Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs |
시그널 제너레이터: 20 V peak-to-valley of 10 ㎐ |
소재지-1 |
| IEC 60601-2-27:2011 |
Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정전류: 30 A 측정거리: 1 500 ㎜ 측정시간: 1/100 s 측정에너지: 600 J |
소재지-1 |
| IEC 60601-2-28:2017 |
Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis |
측정 시간: 1/100 s 측정 저항: 100 ㏀ |
소재지-1 |
| IEC 60601-2-34:2011 |
Medical electrical equipment - Part 2-34: Particular requrements for the safety, including essential performance, of direct blood pressure monitoring equipment <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems |
측정 전압: 5 ㎸ d.c. 측정 소음 레벨: 94 ㏈ |
소재지-1 |
| IEC 60601-2-36:2014 |
Medical electrical equipment - Part 2-36: Particular requirements for the basic safety and essential performance of equipment for extracorporeally induced lithotripsy <제외항목> 201.11.7 Biocompatibility of ME EQUIPMENT and ME SYSTEMS |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 50 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-37:2007+A1:2015 |
Medical electrical equipment - Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
측정 온도: 60 ℃ 측정 주파수: (0.03 ~ 20) ㎒ 측정 출력 세기: 100 W |
소재지-1 |
| IEC 60601-2-4:2010+A1:2018 |
Medical electrical equipment - Part 2-4: Particular requirements for the basic safety and essential performance of cardiac defibrillators <제외항목> 201.11.6.5 Ingress of water of particulate matter into ME Equipment and ME Systems 201.11.6.7 Sterilization of ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 주파수: 500 ㎒ 측정 전력: 10 ㎾ 측정 거리: 1 500 ㎜ 측정 온도: 200 ℃ 측정 저항: 100 ㏀ 측정 에너지: 600 J 인가 힘: 5 000 N 측정 시간: 1/100 s |
소재지-1 |
| IEC 60601-2-43:2010+A1:2017 |
Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 40 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-43:2022 |
Medical electrical equipment - Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 40 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-44:2009+A1:2012+A2:2016 |
Medical electrical equipment - Part 2-44: Particular requirements for basic safety and essential performance of X-ray equipment for computed tomography |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 시간: 1/100 s 측정 각도: 40° 측정 무게: 150 ㎏ 측정 관전압: 50 kVp ~ 150 kVp |
소재지-1 |
| IEC 60601-2-45:2011+A1:2015 |
Medical electrical equipment - Part 2-45: Particular requirements for basic safety and essential performance of mammographic X-ray equipment and mammographic stereotactic devices <제외항목> 201.9.2.101.3 Biopsy needle positioning accuracy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 시간: 1/100 s 측정 각도: 40° 측정 무게: 150 ㎏ 측정 관전압: 20 kVp ~ 49 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-45:2011+A1:2015+A2:2022 |
Medical electrical equipment - Part 2-45: Particular requirements for basic safety and essential performance of mammographic X-ray equipment and mammographic stereotactic devices <제외항목> 201.9.2.101.3 Biopsy needle positioning accuracy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 저항: 100 ㏀ 인가 힘: 5 000 N 측정 시간: 1/100 s 측정 각도: 40° 측정 무게: 150 ㎏ 측정 관전압: 20 kVp ~ 49 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-47:2012 |
Medical electrical equipment - Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems |
주파수 제너레이터: 20 V peak-to-valley of 10 ㎐ |
소재지-1 |
| IEC 60601-2-5:2009 |
Medical electrical equipment - Part 2-5: Particular requirements for the basic safety and essential performance of ultrasonic therapy equipment <제외항목> 201.11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
측정 온도: 60 ℃ 측정 주파수: (0.5 ~ 5) ㎒ 측정 출력: 30 W |
소재지-1 |
| IEC 60601-2-52:2009+A1:2015 |
Medical electrical equipment - Part 2-52: Particular requirements for the basic safety and essential performance of medical beds <제외항목> 201.11.6.5.101 Additional requirements for ingress of water of particulate matter into ME EQUIPMENT and ME SYSTEM |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 측정 커패시턴스: 10 F 측정 에너지: 600 J |
소재지-1 |
| IEC 60601-2-54:2009+A1:2015+A2:2018 |
Medical electrical equipment - Part 2-54: Particular requirements for basic safety and essential performance of X-ray equipment for radiography and radioscopy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 50 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-54:2022 |
Medical electrical equipment - Part 2-54: Particular requirements for basic safety and essential performance of X-ray equipment for radiography and radioscopy |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 50 kVp ~ 150 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ |
소재지-1 |
| IEC 60601-2-57:2011 |
Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 커패시턴스: 10 F 측정 에너지: 600 J 측정 파장: (365 ~ 1 100) ㎚ |
소재지-1 |
| IEC 60601-2-57:2023 |
Medical electrical equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 커패시턴스: 10 F 측정 에너지: 600 J 측정 파장: (365 ~ 1 100) ㎚ |
소재지-1 |
| IEC 60601-2-62:2013 |
Medical electrical equipment - Part 2-62: Particular requirements for the basic safety and essential performance of high intensity therapeutic ultrasound (HITU) equipment <제외항목> 11.6.5 Ingress of water or particulate matter into ME Equipment and ME Systems |
측정 온도: 60 ℃ 측정 주파수: (0.03 ~ 20) ㎒ 측정 출력 세기: 100 W |
소재지-1 |
| IEC 60601-2-63:2012+A1:2017 |
Medical electrical equipment - Part 2-63: Particular requirements for basic safety and essential performance of dental extra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7 m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| IEC 60601-2-63:2012+A1:2017+A2:2021 |
Medical electrical equipment - Part 2-63: Particular requirements for basic safety and essential performance of dental extra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| IEC 60601-2-65:2012+A1:2017 |
Medical electrical equipment - Part 2-65: Particular requirements for basic safety and essential performance of dental intra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5 µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7 m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| IEC 60601-2-65:2012+A1:2017+A2:2021 |
Medical electrical equipment - Part 2-65: Particular requirements for basic safety and essential performance of dental intra-oral X-ray equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 저항: 100 ㏀ 측정 시간: 1/100 s 측정 관전압: 23 kVp ~ 145 kVp 측정 방사선량: 5µ㏉ ~ 8 191 µ㏉ 측정 적산 방사선량: 7m㏉/h ~ 26.8 m㏉/h |
소재지-1 |
| IEC 62304:2006+A1:2015 |
Medical device software. Software life-cycle processes |
- |
소재지-1 |
| IEC 62366:2007+A1:2014 |
Medical devices. Application of usability engineering to medical devices |
- |
소재지-1 |
| IEC 62366-1:2015 |
Medical devices - Part 1: Application of usability engineering to medical devices |
- |
소재지-1 |
| IEC 62366-1:2015+A1:2020 |
Medical devices - Part 1: Application of usability engineering to medical devices |
- |
소재지-1 |
| IEC 80601-2-30:2018 |
Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers |
측정 압력: 330 ㎜Hg |
소재지-1 |
| IEC 80601-2-49:2018 |
Medical electrical equipment - Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment |
1. 인가 전압: 300 Va.c. 2. 측정 전압: 600 Va.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: (- 20~150) ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| IEC 80601-2-60:2019 |
Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 커패시턴스: 10 nF 측정 에너지: 600 J |
소재지-1 |
| ISO 80601-2-56:2017+A1:2018 |
Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement <제외항목> 201.101.3 Time response for a DIRECT MODE CLINICAL THERMOMETER that is not an ADJUSTED MODE CLINICAL THERMOMETER 201.11.7 Biocompatibility of ME EQUIPMENT and ME SYSTEMS |
측정 온도: 50 ℃ |
소재지-1 |
| ISO 80601-2-61:2017 |
Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment <제외항목> 201.11.6.5.101 Additional requirements for ingress of water of particulate matter into ME EQUIPMENT and ME SYSTEM 201.15.3.5.101 Shock and vibration |
인가 전압: 300 V a.c. 측정 전압: 600 V a.c. 측정 전류: 30 A 측정 거리: 1 500 ㎜ 측정 시간: 1/100 s 측정 소음: 94 ㏈ 측정 저항: 100 ㏀ 커패시턴스: 10 nF 측정 에너지: 600 J 측정 정확도: (91~100) % ± (3 counts + EUT 정확도) (81~90) % ± (5 counts + EUT 정확도) (71~80) % ± (7 counts EUT 정확도) |
소재지-1 |
| ISO 80601-2-72:2015 |
Medical electrical equipment - Part 2-72: Particular requirements for basic safety and essential performance of home healthcare environment ventilators for ventilator-dependent patients <제외항목> 201.11.6.7 Sterilization of ME EQUIPMENT or ME SYSTEM 201.12.1 Accuracy of controls and instruments |
1. 인가 전압: 300 Va.c. 2. 측정 전압: 600 Va.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: (- 20~150) ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| ISO 80601-2-74:2017 |
Medical electrical equipment - Part 2-74: Particular requirements for basic safety and essential performance of respiratory humidifying equipment <제외항목> 201.11.6.7 Sterilization of ME EQUIPMENT or ME SYSTEM 201.11.7 Biocompatibility of ME EQUIPMENT and ME SYSTEMS 201.12.1 Accuracy of controls and instruments |
1. 인가 전압: 300 Va.c. 2. 측정 전압: 600 Va.c. 3. 인가 전류: 30 A 4. 측정 전류: 30 A 5. 인가 주파수: 50/60 ㎐ 6. 측정 주파수: 500 ㎒ 7. 측정 전력: 10 ㎾ 8. 측정 거리: 1 500 ㎜ 9. 인가 온도: (- 20~150) ℃ 10. 측정 온도: 200 ℃ 11. 측정 절연저항: ∞ Ω 12. 측정 저항: 100 ㏀ 13. 측정 커패시턴스: 10 nF 14. 인가 에너지: 0.5 J / 5 J 15. 측정 토오크: 1.96 N⋅m 16. 측정 시간: 1/100 s 17. 인가 습도: 95 % R.H. 18. 측정 습도: 70 % R.H. 19. 난연 인가온도: 960 ℃ 20. 측정 각도: 40° 21. 측정 무게: 150 ㎏ 22. 인가 힘: 5 000 N |
소재지-1 |
| ▶ 유/무선 통신기기 |
| ANSI C63.10-2013 |
American National Standard of Procedures for Compliance Testing of Unlicensed Wireless Devices |
9 ㎑ to 40 ㎓ |
소재지-2 |
| ANSI C63.4-2014 |
American National Standard for Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 ㎑ to 40 ㎓ |
9 ㎑ to 40 ㎓ |
소재지-2 |
| ANSI/TIA/EIA-603-E:2016 |
Land Mobile FM or PM Communications Equipment Measurement and Performance Standards |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ARIB STD-T63:2016 |
IMT-2000 DS-CDMA and TDD-CDMA System |
9 ㎑ ~ 26.5 ㎓ |
소재지-2 |
| ARIB STD-T66:2014 |
Second Generation Low Power Data Communication System/ Wireless LAN System |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ARIB STD-T67(1.3):2007 |
Telemeter, Telecontrol Land Data Transmission Radio Equipment For Specified Low-Power Radio Station |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ARIB STD-T71:2014 |
Broadband Mobile Access Communication System (CSMA) |
9 ㎑ ~ 26.5 ㎓ |
소재지-2 |
| AS/NZS 4268:2017 |
Radio equipment and systems - Short range devices - Limits and methods of measurement |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| AS/NZS 4268:2017 Amd 1:2021 |
Radio equipment and systems - Short range devices - Limits and methods of measurement |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| EN 50663:2017 |
Generic standard for assessment of low power electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (10 ㎒ – 300 ㎓) |
10 ㎒ ~ 40 ㎓ |
소재지-2 |
| EN 62311:2008 |
Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields (0 ㎐ ~ 300 ㎓) |
SAR: 30 ㎒ ~ 6 ㎓ MPE: 0 ㎐ ~ 40 ㎓ |
소재지-2 |
| EN 62479:2010 |
Assessment of the Compliance of low Power Electronic and Electrical Equipment With the Basic Restrictions Related to Human Exposure to Electromagnetic Fields (10 ㎒ to 300 ㎓) |
10 ㎒ ~ 40 ㎓ |
소재지-2 |
| EN IEC 62311:2020 |
Assessment of electronic and electrical equipment related to human exposure restrictions for electromagnetic fields(0 ㎐ to 300 ㎓) |
SAR: 30 ㎒ ~ 6 ㎓ MPE: 0 ㎐ ~ 40 ㎓ |
소재지-2 |
| ETSI EN 300 220-1:2017 |
Short Range Devices (SRD) operating in the frequency range 25 ㎒ to 1 000 ㎒; Part 1: Technical characteristics and methods of measurement |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 300 220-2:2017 |
Short Range Devices (SRD) operating in the frequency range 25 ㎒ to 1 000 ㎒; Part 2: Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU for non specific radio equipment |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 300 220-2:2018 |
Short Range Devices (SRD) operating in the frequency range 25 ㎒ to 1 000 ㎒; Part 2: Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU for non specific radio equipment |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 300 328:2019 |
Wideband transmission systems; Data transmission equipment operating in the 2,4 ㎓ band; Harmonised Standard for access to radio spectrum |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 300 330:2017 |
Short Range Devices (SRD); Radio equipment in the frequency range 9 ㎑ to 25 ㎒ and inductive loop systems in the frequency range 9 ㎑ to 30 ㎒; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 300 440:2017 |
Short Range Devices ; Radio equipment to be used in the 1 ㎓ to 40 ㎓ frequency range ; Part 2 |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| ETSI EN 301 091-1:2017 |
Short Range Devices; Transport and Traffic Telematics (TTT); Radar equipment operating in the 76 ㎓ to 77 ㎓ range; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU; Part 1: Ground based vehicular radar |
30 ㎒ ~ 260 ㎓ |
소재지-2 |
| ETSI EN 301 357:2017 |
Codeless audio devices in the range 25 ㎒ to 2000 ㎒ |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 301 441:2016 |
Satellite Earth Stations and Systems (SES); Harmonised Standard for Mobile Earth Stations (MES), including handheld earth stations, for Satellite Personal Communications Networks (S-PCN) operating in the 1,6 ㎓/2,4 ㎓ frequency band under the Mobile Satellite Service (MSS) covering the essential requirements of article 3.2 of the Directive 2014/53/EU |
100 ㎑ ~ 12.75 ㎓ |
소재지-2 |
| ETSI EN 301 511:2017 |
Global System for Mobile communications (GSM); Mobile Stations (MS) equipment; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
30 ㎒ ~ 4 ㎓ |
소재지-2 |
| ETSI EN 301 893:2017 |
Broadband Radio Access Networks (BRAN) ; 5 ㎓ high performance RLAN |
9 ㎑ ~ 26.5 ㎓ |
소재지-2 |
| ETSI EN 301 908-1:2016 |
IMT cellular networks; Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 1: Introduction and common requirements |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 301 908-13:2016 |
IMT cellular networks; Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 13: Evolved Universal Terrestrial Radio Access (E-UTRA) User Equipment (UE) |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 301 908-2:2016 |
IMT cellular networks; Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 2: CDMA Direct Spread (UTRA FDD) User Equipment (UE) |
Test specification for 3G Mobile R&TTE excluding RSE |
소재지-2 |
| ETSI EN 302 065-1:2016 |
Short Range Devices (SRD) using Ultra Wide Band technology (UWB); Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 1: Requirements for Generic UWB applications |
30 ㎒ ~ 40 ㎓ |
소재지-2 |
| ETSI EN 302 065-2:2016 |
Short Range Devices (SRD) using Ultra Wide Band technology (UWB); Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 2: Requirements for UWB location tracking |
30 ㎒ ~ 40 ㎓ |
소재지-2 |
| ETSI EN 302 065-3:2016 |
Short Range Devices (SRD) using Ultra Wide Band technology (UWB); Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 3: Requirements for UWB devices for ground based vehicular applications |
30 ㎒ ~ 40 ㎓ |
소재지-2 |
| ETSI EN 302 065-4:2016 |
Short Range Devices (SRD) using Ultra Wide Band technology (UWB); Harmonised Standard covering the essential requirements of article 3.2 of the Directive 2014/53/EU; Part 4: Material Sensing devices using UWB technology below 10,6 ㎓ |
30 ㎒ ~ 40 ㎓ |
소재지-2 |
| ETSI EN 302 264:2017 |
Short Range Devices; Transport and Traffic Telematics (TTT); Short Range Radar equipment operating in the 77 ㎓ to 81 ㎓ band; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
30 ㎒ ~ 260 ㎓ |
소재지-2 |
| ETSI EN 302 502:2008 |
Broadband Radio Access Networks (BRAN); 5,8 ㎓ fixed broadband data transmitting systems; Harmonized EN covering the essential requirements of article 3.2 of the R&TTE Directive |
30 ㎒ ~ 26.5 ㎓ |
소재지-2 |
| ETSI EN 303 413:2017 |
Satellite Earth Stations and Systems (SES); Global Navigation Satellite System (GNSS) receivers; Radio equipment operating in the 1 164 ㎒ to 1 300 ㎒ and 1 559 ㎒ to 1 610 ㎒ frequency bands; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
30 ㎒ ~ 8.3 ㎓ |
소재지-2 |
| ETSI EN 303 413:2021 |
Satellite Earth Stations and Systems (SES); Global Navigation Satellite System (GNSS) receivers; Radio equipment operating in the 1 164 ㎒ to 1 300 ㎒ and 1 559 ㎒ to 1 610 ㎒ frequency bands; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
30 ㎒ ~ 8.3 ㎓ |
소재지-2 |
| ETSI EN 303 417:2017 |
Wireless power transmission systems, using technologies other than radio frequency beam in the 19 - 21 ㎑, 59 - 61 ㎑, 79 - 90 ㎑, 100 - 300 ㎑, 6 765 - 6 795 ㎑ ranges; Harmonised Standard covering the essential requirements of article 3.2 of Directive 2014/53/EU |
9 ㎑ ~ 18 ㎓ |
소재지-2 |
| ETSI EN 303 687:2023 |
6 ㎓ WAS/RLAN; Harmonised Standard for access to radio spectrumtion techniques |
9 ㎑ ~ 26 ㎓ |
소재지-2 |
| FCC Part 101:2020 |
Fixed microwave services |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 15:2020 |
Radio Frequency Devices |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 2.1091:2019 |
Radiofrequency radiation exposure evaluation; mobile devices |
3 ㎑ ~ 100 ㎓ |
소재지-2 |
| FCC Part 2.1093:2019 |
Radiofrequency radiation exposure evaluation; portable devices |
100 ㎒ ~ 6 ㎓ |
소재지-2 |
| FCC Part 2:2020 |
Frequency allocations and radio treaty matters ; general rules and regulations |
3 ㎑ ~ 100 ㎓ |
소재지-2 |
| FCC Part 22:2020 |
Public mobile service |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 24:2020 |
Personal communications service |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 27:2020 |
Miscellaneous wireless communications service |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 74:2020 |
Experimental Radio, Auxiliary, Special Broadcast and other Program other Distributional Service |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 90:2020 |
Private land mobile radio service |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| FCC Part 95:2020 |
Personal radio services |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| IEEE Std C95.3™:2002 |
IEEE Recommended Practice for Measurements and Computations of Radio Frequency Electromagnetic Fields With Respect to Human Exposure to Such Fields, 100 ㎑ - 300 ㎓ |
300 ㎒ ~ 2.6 ㎓ |
소재지-2 |
| RSS-102 Issue 5:2015 |
Radio Frequency (RF) Exposure Compliance of Radiocommunication Apparatus (All Frequency Band) |
SAR: 100 ㎒ ~ 5 960 ㎒ MPE: 3 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-130:2019 |
Mobile Broadband Service (MBS) Equipment Operating in the Frequency Bands 698 - 756 ㎒ and 777 - 787 ㎒ |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-132:2013 |
Cellular Telephone Systems Operating in the Bands 824 - 849 ㎒ and 869 - 894 ㎒ |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-133:2018 |
2 ㎓ Personal Communications Service |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-139:2015 |
Advanced Wireless Service Equipment Operating in the Bands 1710 - 1780 ㎒ and 2110 - 2180 ㎒ |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-199:2016 |
Broadband Radio Service (BRS) Equipment Operating in the Band 2500 - 2690 ㎒ |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-210:2019 |
Licence-exempt Radio Apparatus (All Frequency Bands) ; Category Ⅰ Equipment |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-216:2016 |
Wireless Power Transfer Devices (Wireless Chargers) |
9 ㎑ ~ 1 ㎓ |
소재지-2 |
| RSS-216:Amendment1(2020) |
Wireless Power Transfer Devices (Wireless Chargers) |
9 ㎑ ~ 1 ㎓ |
소재지-2 |
| RSS-247:2017 |
Digital Transmission Systems (DTSs), Frequency Hopping Systems (FHSs) and Licence-Exempt Local Area Network (LE-LAN) Devices |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| RSS-GEN:2019 |
General Requirements for Compliance of Radio Apparatus |
9 ㎑ ~ 40 ㎓ |
소재지-2 |
| ▶ 전자기적합성 |
| ANSI C63.4:2014 |
American National Standard for Methods of Measurement if Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9㎑ to 40㎓ |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 24 000 ㎒ |
소재지-3 |
| ANSI C63.4a-2017 |
American National Standard for Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 ㎑ to 40 ㎓ - Amendment 1: Test Site Validation |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 000 ㎒ |
소재지-3 |
| AS 61000.6.4:2020 |
Electromagnetic compatibility (EMC), Part 6.4: Generic standards - Emission standard for industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| AS/NZS 61000.6.3:2012 |
Electromagnetic Compatibility(EMC) - Part 6.3: Generic standards - Emission standard for residential, commercial and light - industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A |
소재지-3 |
| AS/NZS CISPR 11:2017 |
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement <제외항목> Table 10–Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site : Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ |
소재지-3 |
| AS/NZS CISPR 14.1:2018 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| AS/NZS CISPR 15:2011 |
Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment <제외항목> 4.2 Insertion loss 4.4 Table 3a-Radiated disturbance limits in the frequency range 9 ㎑ to 30 ㎒ |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 300 ㎒ |
소재지-3 |
| AS/NZS CISPR 32:2015/A1:2020 |
Electromagnetic compatibility of multimedia equipment - Emission requirements |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-3 |
| CISPR 11:2015/A1:2016/A2:2019 |
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement <제외항목> Table 10–Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site : Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ |
소재지-3 |
| CISPR 14-1:2016 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 1: Emission <제외항목> 4.3.2 Table 4-Limits of the magnetic field induced currrent |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| CISPR 14-1:2020 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 1: Emission <제외항목> 4.3.2 Table 4-Limits of the magnetic field induced currrent |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| CISPR 14-2:2015 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity - Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3 V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100 % |
소재지-3 |
| CISPR 14-2:2020 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity -Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100 % |
소재지-3 |
| CISPR 15:2018 |
Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment <제외항목> 4.2 Insertion loss 4.4 Table 3a-Radiated disturbance limits in the frequency range 9 ㎑ to 30 ㎒ |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| CISPR 16-1-2:2014/A1:2017 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 1-2: Radio disturbance and immunity measuring apparatus - Coupling devices for conducted disturbance measurements |
AMN: 9 ㎑ ~ 30 ㎒ CDN: 150 ㎑ ~ 230 ㎒ |
소재지-3 |
| CISPR 16-1-3:2004/A2:2020 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 1-3: Radio disturbance and immunity measuring apparatus - Ancillary equipment Disturbance power |
Absorbing Clamp: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| CISPR 16-1-3:2016 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 1-3: Radio disturbance and immunity measuring apparatus - Ancillary equipment Disturbance power |
Absorbing Clamp: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| CISPR 16-2-1:2014/A1:2017 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-1: Methods of measurement of disturbances and immunity - Conducted disturbance measurements |
9 ㎑ to 30 ㎒ |
소재지-3 |
| CISPR 16-2-2:2010 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-2: Methods of measurement of disturbances and immunity - Measurement of disturbance power |
RFI: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| CISPR 16-2-3:2016/A1:2019 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-3: Methods of measurement of disturbances and immunity - Radiated disturbance measurements |
9 ㎑ ~ 18 ㎓ |
소재지-3 |
| CISPR 16-2-4:2003 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-4: Methods of measurement of disturbances and immunity - Immunity measurements |
CS: 0.15 ㎒ ~ 230 ㎒ RS: 80 ㎒ ~ 3 000 ㎒ |
소재지-3 |
| CISPR 16-2-5:2008 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-5: In situ measurements of disturbing emissions produced by physically large equipment |
Measurements Conditions |
소재지-3 |
| CISPR 32:2015 |
Electromagnetic compatibility of multimedia equipment – Emission Requirement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-3 |
| CISPR 32:2015/A1:2019 |
Electromagnetic compatibility of multimedia equipment – Emission Requirement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-3 |
| CISPR 35:2016 |
Electromagnetic compatibility of multimedia equipment - Immunity requirements |
ESD: 8 ㎸, MF: 1 A/m RS: 3V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 4 ㎸ CS: 3V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| DNV STANDARD FOR CERTIFICATION No.2.4:2016 |
Environmental test specification for electrical, electronic and programmable equipment and systems 4. External power supply failure 5. Power supply variations 14. Electromagnetic compatibilty(EMC) |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 6 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: 0% ~ 100% Conducted low frequency: 50 ㎐ ~ 12 ㎑ |
소재지-3 |
| EN 50121-2:2017 |
Railway applications - Electromagnetic compatibility - Part 2: Emission of the whole railway system to the outside world |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-3 |
| EN 50121-3-1:2017/A1:2019 |
Railway applications - Electromagnetic compatibility - Part 3-1: Rolling stock - Train and complete vehicle |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-3 |
| EN 50121-3-2:2016/A1:2019 |
Railway applications - Electromagnetic compatibility - Part 3-2: Rolling stock - Apparatus |
RE: 30 ㎒ ~ 6 ㎓ CE: 150 ㎑ ~ 30 ㎒ ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 6 ㎓ EFT: 2 ㎸, SURGE: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ |
소재지-3 |
| EN 50121-4:2016/A1:2019 |
Railway applications - Electromagnetic compatibility - Part 4: Emission and immunity of the signalling and telecommunications apparatus |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 1 ㎓ ESD: 8 ㎸, PMF: 300 A/m EFT: 2 ㎸, SURGE: 2 ㎸ MF: 100 A/m ~ 300 A/m CS: 10 V, 0.15 ㎒ ~ 80 ㎒ RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| EN 50130-4:2011/A1:2014 |
Alarm systems - Part 4: Electromagnetic compatibility - Product family standard: Immunity requirements for components of fire, intruder hold up, CCTV, access control and social alarm systems |
ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2.7 ㎓ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 100 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| EN 50155:2017 |
Railway applications- Electronic equipment used rolling stock 5.1 and 13.4.3 Power supply 4.3.6 and 13.4.8 Electromagnetic compatibility tests |
O.V: AC/DC 300 V RE: 30 ㎒ ~ 6 ㎓ CE: 150 ㎑ ~ 30 ㎒ ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 6 ㎓ EFT: 2 ㎸, SURGE: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 40 % Variation (1.5 s) V 100 % Interruption (10 or 20 ms) |
소재지-3 |
| EN 55011:2016/A1:2017 |
Industrial, scientific and medical equipment — Radio-frequency disturbance characteristics — Limits and methods of measurement <제외항목> Table 10–Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site : Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009㎒ ~ 18 ㎓ |
소재지-3 |
| EN 55011:2016/A1:2017/A11:2020 |
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009㎒ ~ 18 ㎓ |
소재지-3 |
| EN 55013:2016 |
Sound and television broadcast receivers and associated equipment - Radio disturbance characteristics - Limits and methods of measurement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ RFI: 30 ㎒ ~ 300 ㎒ ERP: 0.9 ㎓ ~ 18 ㎓ |
소재지-3 |
| EN 55014-1:2017 |
Electromagnetic compatibility. Requirements for household appliances, electric tools and similar apparatus. Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| EN 55014-1:2017/A11:2020 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 1: Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-3 |
| EN 55014-2:2015 |
Requirements for household appliances, electric tools and similar apparatus - Immunity |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3 V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100% |
소재지-3 |
| EN 55014-2:2021 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity - Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3 V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100% |
소재지-3 |
| EN 55015:2019/A11:2020 |
Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 300 ㎒ |
소재지-3 |
| EN 55032:2015 |
Electromagnetic compatibility of multimedia equipment – Emission Requirement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-3 |
| EN 55032:2015/A11:2020 |
Electromagnetic compatibility of multimedia equipment - Emission Requirements |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-3 |
| EN 55035:2017 |
Electromagnetic compatibility of multimedia equipment. Immunity requirements |
ESD: 8 ㎸, MF: 1 A/m RS: 3V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 4 ㎸ CS: 3V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| EN 55035:2017/A11:2020 |
Electromagnetic compatibility of multimedia equipment - Immunity requirements |
ESD: 8 ㎸, MF: 3 A/m RS: 3V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 4 ㎸ CS: 3V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| EN 60601-1-2:2015 |
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral standard : EMC requirements and tests |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-3 |
| EN 60601-1-2:2015/A1:2021 |
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-3 |
| EN 61000-3-2:2019/A1:2021 |
Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤ 16 A per phase) |
≤ 16 A |
소재지-3 |
| EN 61000-3-3:2013/A1:2019 |
ElectroMagnetic Compatibility(EMC) - Part 3-3: Limitation of voltage change, voltage fluctuations and flicker in public low-voltage supply system, for equipment with rated current <= 16 A per phase and not subject to conditional connection <제외항목> 3 phase connection equipment |
≤16 A |
소재지-3 |
| EN 61000-4-11:2004/A1:2017 |
Electromagnetic compatibility (EMC). Testing and measurement techniques. Voltage dips, short interruptions and voltage variations immunity tests |
0 % ~ 100 % |
소재지-3 |
| EN 61000-4-11:2020 |
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase |
0 % ~ 100 % |
소재지-3 |
| EN 61000-4-2:2009 |
Electromagnetic Compatibility(EMC) - Part 4-2: Testing and measurement techniques - Electromagnetic discharge immunity test |
15 ㎸ |
소재지-3 |
| EN 61000-4-3:2006/A2:2010 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| EN 61000-4-3:2020 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| EN 61000-4-4:2012 |
Electromagnetic Compatibility(EMC) - Part 4-4: Testing and measurement techniques - Electrical fast transient/burst immunity test |
± 4 ㎸ |
소재지-3 |
| EN 61000-4-5:2014/A1:2017 |
Amendment 1 - Electromagnetic compatibility (EMC). - Part 4-5: Testing and measurement techniques - Surge immunity test |
4.0 ㎸ |
소재지-3 |
| EN 61000-4-6:2014/AC:2015 |
Electromagnetic Compatibility(EMC) - Part 4-6: Testing and measurement techniques - Immunity to conducted disturbances, induced by radio -frequency fields |
0.15 ㎒ ~ 230 ㎒, 10 V |
소재지-3 |
| EN 61000-4-8:2010 |
Electromagnetic Compatibility(EMC) - Part 4-8: Testing and measurement techniques - Power frequency magnetic field immunity test |
100 A/m |
소재지-3 |
| EN 61000-4-9:2016 |
Electromagnetic Compatibility(EMC) - Part 4-9: Testing and measurement techniques - Pulse magnetic field immunity test |
100 A/m |
소재지-3 |
| EN IEC 61000-6-1:2019 |
Electromagnetic compatibility (EMC) - Part 6-1: Generic standards - Immunity standard for residential, commercial and light-industrial environments |
ESD: 8 ㎸, MS: 3 A/m RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒, 1.4 ㎓ ~ 6 ㎓ EFT/B: 1 ㎸,,Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| EN IEC 61000-6-2:2019 |
Electromagnetic compatibility (EMC) - Part 6-2: Generic standards - Immunity standard for industrial environments |
ESD: 8 ㎸, MS: 30 A/m RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒, 1.4 ㎓ ~ 6 ㎓ EFT/B: 2 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| EN IEC 61000-6-3:2021 |
Electromagnetic compatibility (EMC) - Part 6-3: Generic standards - Emission standard for equipment in residential environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A |
소재지-3 |
| EN IEC 61000-6-4:2019 |
Electromagnetic compatibility (EMC) - Part 6-4: Generic standards - Emission standard for industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| EN 61204-3:2018 |
Low-voltage switch mode power supplies - Part 3: Electromagnetic compatibility (EMC) |
CE: 0.15 ㎒ ~ 30 ㎒ RE, RFI: 30 ㎒ ~ 1 000 ㎒ F/H: ≤16 A ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 2 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| EN 61326-3-1:2017 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-1: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - General industrial applications |
ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 3 ㎸ Surge: 4 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-3 |
| EN 61326-3-2:2018 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-2: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - Industrial applications with specified electromagnetic environment |
ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 2 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MS: 100 A/m Dips: 30 % ~ 100 % |
소재지-3 |
| EN 61547:2009 |
Equipment for general light purposes - EMC immunity requirements |
ESD: 4 ㎸ ~ 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: (0 ~ 100) % |
소재지-3 |
| EN 62040-2:2018 |
Uninterruptible power systems (UPS) - Part 2: Electromagnetic compatibility (EMC) requirements |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 1 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m |
소재지-3 |
| EN IEC 61000-3-2:2019 |
ElectroMagnetic Compatibility(EMC) - Part 3-2: Limitation for harmonics current emissions (equipment input current <= 16 A per phase) <제외항목> 3 phase connection equipment |
≤16 A |
소재지-3 |
| EPRI TR-102323:1997 |
Guidelines for Electromagnetic Interference Testing in Power Plants B-3.3 Surge B-3.4 Fast transient and Impulse tests B-3.5 Electrostatic discharge (ESD) |
Surge: 3 ㎸ ESD: 15 ㎸ EFT: 3 ㎸ |
소재지-3 |
| ETSI EN 300 386 V2.1.1:2016 |
Telecommunication network equipment; ElectroMagnetic Compatibility (EMC) requirements |
RE: 30 ㎒ ~ 1 ㎓ CE: 150 ㎑ ~ 30 ㎒ ESD: 8 ㎸, RS: 10 V/m 80 ㎒ ~ 2 700 ㎒ EFT: 1 ㎸, SURGE: 2 ㎸ CS: 3 V V-dip: 100 %, 30 % H/F: ≤16 A |
소재지-3 |
| ETSI EN 301 489-1 V2.2.3:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-13:2002 |
Electromagnetic compatibility and Radio spectrum Matters (ERM); ElectroMagnetic Compatibility(EMC) standard for radio equipment and services; Part 13: Specific conditions for Citizens Band (CB) radio and ancillary equipment (speech and non-speech) |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-15 V2.2.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 15: Specific conditions for commercially available amateur radio equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-17 V3.2.4:2020 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 17: Specific conditions for Broadband Data Transmission Systems |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-19 V2.2.1:2022 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 19: Specific conditions for Receive Only Mobile Earth Stations (ROMES) operating in the 1,5 ㎓ band providing data communications and GNSS receivers operating in the RNSS band providing positioning, navigation, and timing data; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-2 V2.1.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 2: Specific conditions for radio paging equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-2:2002 |
Electromagnetic compatibility and Radio spectrum Matters (ERM); ElectroMagnetic Compatibility(EMC) standard for radio equipment and services; Part 2: Specific conditions for radio paging equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-3 V2.3.2:2023 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 3: Specific conditions for Short Range Devices (SRD) operating on frequencies between 9 ㎑ and 246 ㎓; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-33 V2.2.1(2019-04) |
Electromagnetic Compatibility(EMC) Standard for radio equipment and services ; Part 33:specific conditions for Ultra-Wideband(UWB) devices: Harmonized Standard covering the essential requirements of article 3.1(b) of Directive 2014/53/EU |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % CTI : Pulse 1, 2a, 2b, 3a, 3b, 4 |
소재지-3 |
| ETSI EN 301 489-34 V2.1.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 34: Specific conditions for External Power Supply (EPS) for mobile phones |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-4 V3.3.1:2021 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 4: Specific conditions for fixed radio links and ancillary equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-5 V2.2.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 5: Specific conditions for Private land Mobile Radio (PMR) and ancillary equipment (speech and non-speech) and Terrestrial Trunked Radio (TETRA) |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-51 V2.1.1(2019-04) |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 51: Specific conditions for Automotive, Ground based Vehicles and Surveillance Radar Devices using 24,05 ㎓ to 24,25 ㎓z, 24,05 ㎓ to 24,5 ㎓, 76 ㎓ to 77 ㎓ and 77 ㎓ to 81 ㎓; Harmonised Standard covering the essential requirements of article 3.1(b) of Directive 2014/53/EU |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-52 V1.2.1:2021 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 52: Specific conditions for Cellular Communication User Equipment (UE) radio and ancillary equipment; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| ETSI EN 301 489-9 V2.1.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 9: Specific conditions for wireless microphones, similar Radio Frequency (RF) audio link equipment, cordless audio and in-ear monitoring devices |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| FCC Part 15:2020 |
Radio Frequency Devices <제외항목> 15.115 TV interfaces devices, including cable system terminal devices 15.117 TV Broadcast receivers |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 ㎓ |
소재지-3 |
| FCC Part 18:2020 |
Industrial, Scientific and Medical Equipment <제외항목> Radiated Emission test at 30 m, 300m and 1 600m distance. |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 40 ㎓ |
소재지-3 |
| IACS URE 10 Rev.7:2018 |
International association of classification societies E10 3. External power supply failure 4. Power supply variations 13. Electrostatic discharge 14. Electromagnetic field 15. Conducted Low Frequency 16. Conducted Radio Frequency 17. Burst/Fast Transients 18. Surge/voltage 19. Radiated Emission 20. Conducted Emission |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 6 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: 0% ~ 100% Conducted low frequency: 50 ㎐ ~ 12 ㎑ |
소재지-3 |
| IC RSS-210:2016 |
License-exempt Radio Apparatus(All Frequency Bands) : Category I Equipment |
RE: 0.009 ㎒ ~ 40 ㎓ |
소재지-3 |
| IC RSS-210:2019/A1:2020 |
Licence-Exempt Radio Apparatus: Category I Equipment |
RE: 0.009 ㎒ ~ 40 ㎓ |
소재지-3 |
| ICES-001:2020 |
Industrial, Scientific and Medical (ISM) Equipment |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 1 ㎓ |
소재지-3 |
| ICES-003:2019 |
Information Technology Equipment (Including Digital Apparatus) — Limits and Methods of Measurement |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 000 ㎒ |
소재지-3 |
| IEC 60533:2015 |
Electrical and electronic installations in ships - Electromagnetic compatibility <제외항목> 7.2 Minimum immunity requirements, Table 4 - Minimum immunity requirements for ship equipment, AC and DC power Port, Conducted low frequency interference |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 20 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-3 |
| IEC 60571:2012 |
Railway applications – Electronic equipment used on rolling stock 5.1 Power supply 5.2 and 12.2.6 Supply over voltages 5.4 and 12.2.7 Surge, electrostatic discharge (ESD) and transient burst susceptibility tests 5.5 Electromagnetic compatibility tests 12.2.8 Radio interference test |
O.V: AC/DC 300 V SURGE: 7.4 ㎸, ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 2.5 ㎓ BURST: 4.4 ㎸ RE: 30 ㎒ ~ 1 ㎓ CE: 9 ㎑ ~ 30 ㎒ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 40 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-3 |
| IEC 60601-1-2:2014 |
Medical electrical equipment - Part 1-2: General requirement for safety - Collateral standard : Electromagnetic compatibility Requirements and test |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-3 |
| IEC 60601-1-2:2014/A1:2020 |
General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A ESD: ± 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ Proximity fields : 28 V/m, 385 ㎒ ~ 5 785 ㎒ EFT/B: ± 2 ㎸ Surge: ± 2 ㎸ CS: 0.15 ㎒ ~ 80 ㎒ MF: 30 A/m Dips: 30 % ~ 100 % PMF: 30 ㎑(8 A/m), 134.2 ㎑(65 A/m), 13.56 ㎒(7.5 A/m) |
소재지-3 |
| IEC 60945:2002 |
Maritime navigation and radio communication equipment and systems General requirements 4.3. Power supply 7. Power supply-Methods of testing and required test results 9. Electromagnetic emission 10. Immunity to electromagnetic environment |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 20 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-3 |
| IEC 61000-3-2:2018 |
ElectroMagnetic Compatibility(EMC) - Part 3-2: Limits for harmonic current emissions (equipment input current <= 16 A per phase) <제외항목> 3 phase connection equipment |
≤16 A |
소재지-3 |
| IEC 61000-3-2:2018/A1:2020 |
Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤16 A per phase) <제외항목> 3 phase connection equipment |
≤16 A |
소재지-3 |
| IEC 61000-3-3:2013/A1:2017 |
ElectroMagnetic Compatibility(EMC) - Part 3-3: Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current <= 16 A per phase and not subject to conditional connection <제외항목> 3 phase connection equipment |
≤16 A |
소재지-3 |
| IEC 61000-4-11:2004/A1:2017 |
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests |
0 % ~ 100 % |
소재지-3 |
| IEC 61000-4-11:2020 |
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase |
0 % ~ 100 % |
소재지-3 |
| IEC 61000-4-2:2008 |
ElectroMagnetic Compatibility(EMC) - Part 4-2: Testing and measurement techniques - Electromagnetic discharge immunity test |
15 ㎸ |
소재지-3 |
| IEC 61000-4-3:2006+A1:2007+A2:2010 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| IEC 61000-4-3:2020 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| IEC 61000-4-39:2017 |
Electromagnetic compatibility (EMC) – Part 4-39: Testing and measurement techniques - Radiated fields in close proximity - Immunity test |
9 ㎑ ~ 300 ㎒ / Only used section 6.1 for magnetic field immunity |
소재지-3 |
| IEC 61000-4-4:2012 |
ElectroMagnetic Compatibility(EMC) - Part 4-4: Testing and measurement techniques - Electrical fast transient/burst immunity test |
4.0 ㎸ |
소재지-3 |
| IEC 61000-4-5:2014/A1:2017 |
Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement techniques - Surge immunity test |
4.0 ㎸ |
소재지-3 |
| IEC 61000-4-6:2013 |
ElectroMagnetic Compatibility(EMC) - Part 4-6: Testing and measurement techniques - Immunity to conducted disturbances, induced by radio - frequency fields |
0.15 ㎒ ~ 230 ㎒, 10 V |
소재지-3 |
| IEC 61000-4-8:2009 |
ElectroMagnetic Compatibility(EMC) - Part 4-8: Testing and measurement techniques - Power frequency magnetic field immunity test |
100 A/m |
소재지-3 |
| IEC 61000-4-9:2016 |
Electromagnetic compatibility (EMC) - Part 4-9: Testing and measurement techniques - Impulse magnetic field immunity test |
300 A/m |
소재지-3 |
| IEC 61000-6-1:2016 |
Electromagnetic compatibility (EMC) - Part 6-1: Generic standards - Immunity standard for residential, commercial and light-industrial environments |
ESD: 8 ㎸, MS: 3 A/m RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒, 1.4 ㎓ ~ 6 ㎓ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| IEC 61000-6-2:2016 |
Electromagnetic compatibility (EMC) - Part 6-2: Generic standards - Immunity standard for industrial environments |
ESD: 8 ㎸, MS: 30 A/m RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-3 |
| IEC 61000-6-3:2020 |
Electromagnetic compatibility (EMC) - Part 6-3: Generic standards - Emission standard for equipment in residential environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A |
소재지-3 |
| IEC 61000-6-4:2018 |
Electromagnetic compatibility (EMC) - Part 6-4: Generic standards - Emission standard for industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ |
소재지-3 |
| IEC 61204-3:2016 |
Low voltage power supplies, d.c. output - Part 3: ElectroMagnetic Compatibility(EMC) |
CE: 0.15 ㎒ ~ 30 ㎒ RE, RFI: 30 ㎒ ~ 1 000 ㎒. F/H: ≤16 A, ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 20 % ~ 100 % |
소재지-3 |
| IEC 61326-1:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS: 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF: 30 A/m, Burst: 2 kV, Surge: 2 kV, CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dip: 30 ~ 100 % |
소재지-3 |
| IEC 61326-2-1:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-1: Particular requirements - Test configurations, operational conditions and performance criteria for sensitive test and measurement equipment for EMC unprotected applications |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61326-2-2:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-2: Particular requirements - Test configurations, operational conditions and performance criteria for portable testing, measuring and monitoring equipment used in low-voltage distribution systems |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61326-2-3:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-3: Particular requirements - Test configuration, operational conditions and performance criteria for transducers with integrated or remote signal conditioning |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61326-2-4:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-4: Particular requirements - Test configurations, operational conditions and performance criteria for insulation monitoring devices according to IEC 61557-8 and for equipment for insulation fault location according to IEC 61557-9 |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 10 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61326-2-5:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-5: Particular requirements - Test configurations, operational conditions and performance criteria for field devices with field bus interfaces according to IEC 61784-1 |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61326-2-6:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6 V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-3 |
| IEC 61326-3-1:2017 |
Electrical equipment for measurement, control and laboratory use - EMC requirements – Part 3-1: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) – General industrial applications |
ESD: 8 kV RS : 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 3 kV, Surge : 4 kV, CS : 10 V, Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61326-3-2:2017 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-2: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - Industrial applications with specified electromagnetic environment |
ESD: 8 kV RS : 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 100 A/m, Burst : 2 kV, Surge : 2 kV, CS : 10 V, Dip : 30 ~ 100 % |
소재지-3 |
| IEC 61547:2009 |
Equipment for general lighting purposes - EMC immunity requirements |
ESD: 8 ㎸, Dips: (0 ~ 100) % RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ |
소재지-3 |
| IEC 61547:2020 |
Equipment for general lighting purposes - EMC immunity requirements |
ESD: 8 kV RS : 3 V/m, 80 ㎒ ~ 1 ㎓, MF : 3 A/m, Burst : 1 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-3 |
| IEC 62040-2:2016 |
Uninterruptible power system (UPS) Part 2: ElectroMagnetic Compatibility(EMC) requirements <제외항목> 7.4 Immunity to low-frequency signals |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, MS: 30 A/m RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ |
소재지-3 |
| IEC 62236-2:2018 |
Railway applications - Electromagnetic compatibility - Part 2: Emission of the whole railway system to the outside world |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Burst: 2 ㎸, Surge: 2 ㎸, RS: 20 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| IEC 62236-3-1:2018 |
Railway applications - Electromagnetic compatibility - Part 3-1: Rolling stock - Train and complete vehicle |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-3 |
| IEC 62236-3-2:2018 |
Railway applications - Electromagnetic compatibility - Part 3-2: Rolling stock - Apparatus |
CE: 0.15 ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, CS: 10 V, 0.15 ㎒ ~ 80 ㎒ EFT: 2 ㎸, Surge: 2 ㎸, RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| IEC 62236-4:2018 |
Railway applications - Electromagnetic compatibility Part 4: Emission and immunity of the signalling and telecommunications apparatus |
CE : 0.15 ㎑ ~ 30 ㎒ RE : 30 ㎒ ~ 6 ㎓ ESD : 8 ㎸, SURGE : ± 2 ㎸ RS : 20 V/m, 80 ㎒ ~ 6 ㎓ EFT : 2 ㎸ CS : 10 V, 0.15 ㎒ ~ 80 ㎒ MF : 0, 16.7, 50, 60 ㎐ / 300 A/m |
소재지-3 |
| KRS CS 0003-13 (R) |
철도차량 차상신호장치 시험방법 4.3절 아. 전기자기 적합성 시험 |
CE: 0.15 ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Burst: 2 ㎸, Surge: 2 ㎸, RS: 20 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KRS SG 0015-20 (R) |
전자연동장치 4.7.3 전기자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KRS SG 0033-21 (R) |
유절연가청주파수(AF) 궤도회로 4.2.2절 6) 전기자기적합성 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KRS SG 0038-21 (R) |
무절연가청주파수(AF) 궤도회로 4.5.3 전기자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KRS SG 0054-21 (R) |
단선 자동폐색제어장치 4.2.2절 6) 전자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KRS SG 0055-21 (R) |
복선 자동폐색제어장치 4.2.2절 6) 전자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KRS SG 0067-14 (R) |
지상 신호제어설비 시험방법 4.3.4 전기자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-3 |
| KR:2020 |
제조법 및 형식승인 등에 관한 지침 23. 자동화시스템 3. 전원상실시험 4. 전원변동시험 14. 정전기방전내성시험 15. 방사전자계 내성시험 16. 전도성 저주파 간섭에 대한 내성시험 17. 전도성 고주파 간섭에 대한 내성시험 18. 버스트 내성시험 19. 서지 내성시험 20. 발산방사 측정시험 21. 전도방사 측정시험 |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 6 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: 0% ~ 100% Conducted low frequency: 50 ㎐ ~ 12 ㎑, 1~10% or Max 2 W |
소재지-3 |
| KS C IEC 60571:2012 |
철도 적용 - 철도 차량용 전자기기 5.1 공급 전원 5.2 와 12.2.6 과전압 5.4 와 12.2.7 서지, 정전기와 전기적 빠른 과도현상/버스트 내성시험 5.5 전자기 적합성 12.2.8 무선 장해 |
O.V: AC/DC 300 V ESD: 8 ㎸, SURGE: 7.4 ㎸ RS: 20 V/m, 80 ㎒ ~ 2.5 ㎓ BURST: 4.4 ㎸ RE: 30 ㎒ ~ 1 ㎓ CE: 9 ㎑ ~ 30 ㎒ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 40 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-3 |
| KS X IEC 60945:2002 |
해상 항해, 무선통신 기기 및 시스템 - 일반 요구사항 - 시험 방법과 요구되는 시험결과 4.3. 공급 전원 7. 공급 전원 시헙방법 및 요구되는 결과 9. 전자파 장해 10. 전자파내성 |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 20 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-3 |
| LR Type approval system:2020 |
LR Type approval system - Test specification number1 2013 - Electrical equipment control and monitoring equipment instrumentation and internal communication equipment computer and peripheral equipment 8. Power supply variation test 9. Power supply failure test 22. Immunity to conducted high frequency interference 23. Immunity to conducted low frequency interference 24. Immunity to radiated radio frequency field 25. Immunity to fast low energy transients (bursts) 26. Immunity to slow high energy transients (surges) 27. Immunity to electrostatic discharge (ESD) 29. Radiated emissions 30. Conducted emissions |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: Voltage 20 % Variation (1.5 s) Freq. 10 % Variation (5 s) Voltage 100 % Dip (60 s) Conducted low frequency: 50 ㎐ ~ 12 ㎑, 1 % ~ 10 % or Max 2 W |
소재지-3 |
| MIL-STD-461D:1993 |
Requirements for the control of electromagnetic interference characteristics of subsystems and equipment CE101 CE102 CS101 CS114 CS115 CS116 RE101 RE102 RS101 |
CE: 30 ㎐ ~ 10 ㎒ RE: 30 ㎐ ~ 18 ㎓ CS: 30 ㎐ ~ 400 ㎒ RS: 30 ㎐ ~100 ㎑ |
소재지-3 |
| MIL-STD-461E:1999 |
Requirements for the control of electromagnetic interference characteristics of subsystems and equipment CE101 CE102 CS101 CS106 CS114 CS115 CS116 RE101 RE102 RS101 RS103 |
CE: 30 ㎐ - 10 ㎒ RE: 30 ㎐ - 18 ㎓ CS: 30 ㎐ - 200 ㎒ RS: 30 ㎐ - 18 ㎓, 50 V/m |
소재지-3 |
| MIL-STD-461F:2007 |
Requirements for the control of electro magnetic interference characteristics of subsystems and equipment CE101 CE102 CS101 CS106 CS114 CS115 CS116 RE101 RE102 RS101 RS103 |
CE: 30 ㎐ - 10 ㎒ RE: 30 ㎐ - 18 ㎓ CS: 30 ㎐ - 200 ㎒ RS: 30 ㎐ - 18 ㎓, 50 V/m |
소재지-3 |
| MIL-STD-461G:2015 |
Requirements for the control of electro magnetic interference characteristics of subsystems and equipment CE101 CE102 CS101 CS114 CS115 CS116 CS118 RE101 RE102 RS101 RS103 |
CE: 30 ㎐ ~ 10 ㎒ RE: 30 ㎐ ~ 18 ㎓ CS: 30 ㎐ ~ 200 ㎒ RS: 30 ㎐ ~ 18 ㎓, 50 V/m ESD 15 ㎸ |
소재지-3 |
| RSS-Gen:2014 |
General Requirements and Information for the Certification of Radio Apparatus |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 ㎓ |
소재지-3 |
| RSS-Gen:2018/A1:2019/A2:2021 |
General Requirements for Compliance of Radio Apparatus |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 ㎓ |
소재지-3 |
| TEC/SD/DD/EMC-221/05/OCT-16 |
Electromagnetic compatibility Standard for Telecommunication Equipment |
CE: 0.15㎒~ 30㎒ RE: 30㎒~ 6 000㎒ ESD: 8㎸ RS: 10 V/m, 80㎒~ 6 000㎒ EFT/B: 1㎸, Surge: 2㎸ CS: 3 V, 0.15㎒~ 80㎒ Dips: 30%~ >95% |
소재지-3 |
| U.SNRC RG-1.10:2003 |
Guidelines for evaluating electromagnetic and radio-frequency interference in safety-related instrumentation and control system - Application Below 100 A <제외항목> Table 7 - IEC 61000-4-10, 12, 13, 16 |
CE: 30 ㎐ ~ 10 ㎒ RE: 30 ㎐ ~ 18 ㎓ CS: 30 ㎐ ~ 200 ㎒ RS: 30 ㎐ ~ 18 ㎓ SURGE: 6 ㎸ EFT: 4 ㎸, ESD: 15 ㎸ |
소재지-3 |
| U.SNRC RG-1.180:2003 |
Guidelines for evaluating electromagnetic and radio-frequency interference in safety-related instrumentation and control system - Application Below 100 A <제외항목> Table 7 - IEC 61000-4-10, 12, 13, 16 |
CE: 30 ㎐ ~ 10 ㎒ RE: 30 ㎐ ~ 18 ㎓ CS: 30 ㎐ ~ 200 ㎒ RS: 30 ㎐ ~ 18 ㎓ SURGE: 6 ㎸ EFT: 4 ㎸, ESD: 15 ㎸ |
소재지-3 |
| ANSI C63.4:2014 |
American National Standard for Methods of Measurement if Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9㎑ to 40㎓ |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 000 ㎒ |
소재지-4 |
| ANSI C63.4a-2017 |
American National Standard for Methods of Measurement of Radio-Noise Emissions from Low-Voltage Electrical and Electronic Equipment in the Range of 9 ㎑ to 40 ㎓ - Amendment 1: Test Site Validation |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 000 ㎒ |
소재지-4 |
| AS 61000.6.4:2020 |
Electromagnetic compatibility (EMC), Part 6.4: Generic standards - Emission standard for industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| AS/NZS 61000.6.3:2012 |
Electromagnetic Compatibility(EMC) - Part 6.3: Generic standards - Emission standard for residential, commercial and light - industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A |
소재지-4 |
| AS/NZS CISPR 11:2017 |
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement <제외항목> Table 10–Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site: Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ |
소재지-4 |
| AS/NZS CISPR 14.1:2018 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| AS/NZS CISPR 15:2011 |
Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment <제외항목> 4.2 Insertion loss |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 300 ㎒ |
소재지-4 |
| AS/NZS CISPR 32:2015/A1:2020 |
Electromagnetic compatibility of multimedia equipment - Emission requirements |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-4 |
| CISPR 11:2015/A1:2016/A2:2019 |
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement <제외항목> Table 10–Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site: Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ |
소재지-4 |
| CISPR 14-1:2016 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 1: Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| CISPR 14-1:2020 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 1: Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| CISPR 14-2:2015 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity -Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100 % |
소재지-4 |
| CISPR 14-2:2020 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity -Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100 % |
소재지-4 |
| CISPR 15:2018 |
Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 300 ㎒ |
소재지-4 |
| CISPR 16-1-2:2014/A1:2017 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 1-2: Radio disturbance and immunity measuring apparatus - Coupling devices for conducted disturbance measurements |
AMN: 9 ㎑ ~ 30 ㎒ CDN: 150 ㎑ ~ 230 ㎒ |
소재지-4 |
| CISPR 16-1-3:2004/A2:2020 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 1-3: Radio disturbance and immunity measuring apparatus - Ancillary equipment Disturbance power |
Absorbing Clamp: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| CISPR 16-1-3:2016 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 1-3: Radio disturbance and immunity measuring apparatus - Ancillary equipment Disturbance power |
Absorbing Clamp: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| CISPR 16-2-1:2014/A1:2017 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-1: Methods of measurement of disturbances and immunity - Conducted disturbance measurements |
9 ㎑ to 30 ㎒ |
소재지-4 |
| CISPR 16-2-2:2010 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-2: Methods of measurement of disturbances and immunity - Measurement of disturbance power |
RFI: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| CISPR 16-2-3:2016/A1:2019 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-3: Methods of measurement of disturbances and immunity - Radiated disturbance measurements |
9 ㎑ ~ 18 ㎓ |
소재지-4 |
| CISPR 16-2-4:2003 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-4: Methods of measurement of disturbances and immunity - Immunity measurements <제외항목> Radiated immunity |
CS: 0.15 ㎒ ~ 230 ㎒ RS: 80 ㎒ ~ 3 000 ㎒ |
소재지-4 |
| CISPR 16-2-5:2008 |
Specification for radio disturbance and immunity measuring apparatus and methods - Part 2-5: In situ measurements of disturbing emissions produced by physically large equipment |
Measurements Conditions |
소재지-4 |
| CISPR 32:2015 |
Electromagnetic compatibility of multimedia equipment – Emission Requirement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-4 |
| CISPR 32:2015/A1:2019 |
Electromagnetic compatibility of multimedia equipment – Emission Requirement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-4 |
| CISPR 35:2016 |
Electromagnetic compatibility of multimedia equipment - Immunity requirements |
ESD: 8 ㎸, MF: 1 A/m RS: 3V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 4 ㎸ CS: 3V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| DNV STANDARD FOR CERTIFICATION No.2.4:2016 |
Environmental test specification for electrical, electronic and programmable equipment and systems 4. External power supply failure 5. Power supply variations 14. Electromagnetic compatibilty(EMC) |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 6 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: 0% ~ 100% Conducted low frequency: 50 ㎐ ~ 12 ㎑ |
소재지-4 |
| EN 50121-2:2017 |
Railway applications - Electromagnetic compatibility - Part 2: Emission of the whole railway system to the outside world |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-4 |
| EN 50121-3-1:2017/A1:2019 |
Railway applications - Electromagnetic compatibility - Part 3-1: Rolling stock - Train and complete vehicle |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-4 |
| EN 50121-3-2:2016/A1:2019 |
Railway applications - Electromagnetic compatibility - Part 3-2: Rolling stock - Apparatus |
RE: 30 ㎒ ~ 6 ㎓ CE: 150 ㎑ ~ 30 ㎒ ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 6 ㎓ EFT: 2 ㎸, SURGE: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ |
소재지-4 |
| EN 50121-4:2016/A1:2019 |
Railway applications – Electromagnetic compatibility – Part4: Emission and immunity of the signalling and telecommunications apparatus |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 1 ㎓ ESD: 8 ㎸, PMF: 300 A/m EFT: 2 ㎸, SURGE: 2 ㎸ MF: 100 A/m ~ 300 A/m CS: 10 V, 0.15 ㎒ ~ 80 ㎒ RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| EN 50130-4:2011/A1:2014 |
Alarm systems - Part 4: Electromagnetic compatibility - Product family standard: Immunity requirements for components of fire, intruder hold up, CCTV, access control and social alarm systems |
ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 100 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| EN 50155:2017 |
Railway applications- Electronic equipment used rolling stock 5.1 Power supply 5.2 and 12.2.6 Supply over voltages 5.4 and 12.2.7 Surge, electrostatic discharge (ESD) and transient burst susceptibility tests 5.5 Electromagnetic compatibility tests 12.2.8 Radio interference test |
O.V: AC/DC 300 V RE: 30 ㎒ ~ 6 ㎓ CE: 150 ㎑ ~ 30 ㎒ ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 6 ㎓ EFT: 2 ㎸, SURGE: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 40 % Variation (1.5 s) V 100 % Interruption (10 or 20 ms) |
소재지-4 |
| EN 50500:2008/A1:2015 |
Measurement procedures of magnetic field levels generated by electronic and electrical apparatus in the railway environment with respect to human exposure |
EMF: 0 ㎐ ~ 20 ㎑ |
소재지-4 |
| EN 55011:2016/A1:2017 |
Industrial, scientific and medical equipment – Radio-frequency disturbance characteristics – Limits and methods of measurement <제외항목> Table 10–Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site : Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009㎒ ~ 18 ㎓ |
소재지-4 |
| EN 55011:2016/A1:2017/A11:2020 |
Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics – Limits and methods of measurement <제외항목> Rated input power of > 75 ㎸A in Table 6 Radiated Emission test at 10 m distance. Table 10 – Electromagnetic radiation disturbance limits for class A group 2 equipment measured on a test site: Measuring distance = 30m |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009㎒ ~ 18 ㎓ |
소재지-4 |
| EN 55013:2016 |
Sound and television broadcast receivers and associated equipment - Radio disturbance characteristics - Limits and methods of measurement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ERP: 0.9 ㎓ ~ 18 ㎓ |
소재지-4 |
| EN 55014-1:2017 |
Electromagnetic compatibility. Requirements for household appliances, electric tools and similar apparatus. Emission |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| EN 55014-1:2017/A11:2020 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 1: Emission <제외항목> Radiated Emission test at 10 m distance. |
CE: 0.15 ㎒ ~ 30 ㎒ RFI: 30 ㎒ ~ 300 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ |
소재지-4 |
| EN 55014-2:2015 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity -Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3 V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100% |
소재지-4 |
| EN 55014-2:2021 |
Electromagnetic compatibility - Requirements for household appliances, electric tools and similar apparatus - Part 2: Immunity - Product family standard |
ESD: 8 ㎸ RS: 80 ㎒ ~ 1 000 ㎒, 3 V/m EFT: 1 ㎸, Surge: 2 ㎸ CS: 0.15 ㎒ ~ 230 ㎒, 3V DIP: 30 % ~ 100% |
소재지-4 |
| EN 55015:2013/A1:2015 |
Limits and methods of measurement of radio disturbance characteristics of electrical lights and similar equipment <제외항목> 4.2 Insertion loss |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 300 ㎒ |
소재지-4 |
| EN 55015:2019/A11:2020 |
Limits and methods of measurement of radio disturbance characteristics of electrical lighting and similar equipment |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 300 ㎒ |
소재지-4 |
| EN 55032:2015 |
Electromagnetic compatibility of multimedia equipment - Emission Requirement |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-4 |
| EN 55032:2015/A11:2020 |
Electromagnetic compatibility of multimedia equipment - Emission Requirements |
CE: 0.15 ㎒ ~ 2 150 ㎒ RE: 30 ㎒ ~ 18 000 ㎒ |
소재지-4 |
| EN 55035:2017 |
Electromagnetic compatibility of multimedia equipment. Immunity requirements |
ESD: 8 ㎸, MF: 1 A/m RS: 3V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 4 ㎸ CS: 3V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| EN 55035:2017/A11:2020 |
Electromagnetic compatibility of multimedia equipment - Immunity requirements |
ESD: 8 ㎸, MF: 3 A/m RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 4 ㎸ CS: 3V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| EN 60601-1-2:2015 |
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance -Collateral standard : EMC requirements and tests |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-4 |
| EN 60601-1-2:2015/A1:2021 |
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-4 |
| EN 61000-3-2:2019/A1:2021 |
Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤ 16 A per phase) |
≤ 16 A |
소재지-4 |
| EN 61000-3-3:2013/A1:2019 |
ElectroMagnetic Compatibility(EMC) - Part 3-3: Limitation of voltage change, voltage fluctuations and flicker in public low-voltage supply system, for equipment with rated current <= 16 A per phase and not subject to conditional connection <제외항목> 3 phase connection equipment |
≤16 A |
소재지-4 |
| EN 61000-4-11:2004/A1:2017 |
Electromagnetic compatibility (EMC). Testing and measurement techniques. Voltage dips, short interruptions and voltage variations immunity tests |
0 % ~ 100 % |
소재지-4 |
| EN 61000-4-11:2020 |
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase |
0 % ~ 100 % |
소재지-4 |
| EN 61000-4-2:2009 |
Electromagnetic Compatibility(EMC) - Part 4-2: Testing and measurement techniques - Electromagnetic discharge immunity test |
30 ㎸ |
소재지-4 |
| EN 61000-4-3:2006/A1:2008/A2:2010 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| EN 61000-4-3:2020 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| EN 61000-4-4:2012 |
Electromagnetic Compatibility(EMC) - Part 4-4: Testing and measurement techniques - Electrical fast transient/burst immunity test |
± 4 ㎸ |
소재지-4 |
| EN 61000-4-5:2014/A1:2017 |
Amendment 1 - Electromagnetic compatibility (EMC). - Part 4-5: Testing and measurement techniques - Surge immunity test |
4.0 ㎸ |
소재지-4 |
| EN 61000-4-6:2014/AC:2015 |
Electromagnetic Compatibility(EMC) - Part 4-6: Testing and measurement techniques - Immunity to conducted disturbances, induced by radio - frequency fields |
0.15 ㎒ ~ 230 ㎒, 10 V/m |
소재지-4 |
| EN 61000-4-8:2010 |
Electromagnetic Compatibility(EMC) - Part 4-8: Testing and measurement techniques - Power frequency magnetic field immunity test |
1 000 A/m |
소재지-4 |
| EN 61000-4-9:2016 |
Electromagnetic Compatibility(EMC) - Part 4-9: Testing and measurement techniques - Pulse magnetic field immunity test |
1 000 A/m |
소재지-4 |
| EN IEC 61000-6-1:2019 |
Electromagnetic compatibility (EMC) - Part 6-1: Generic standards - Immunity standard for residential, commercial and light-industrial environments |
ESD: 8 ㎸, MS: 3 A/m RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒, 1.4 ㎓ ~ 6 ㎓ EFT/B: 1 ㎸,,Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| EN IEC 61000-6-2:2019 |
Electromagnetic compatibility (EMC) - Part 6-2: Generic standards - Immunity standard for industrial environments |
ESD: 8 ㎸, MS: 30 A/m RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒, 1.4 ㎓ ~ 6 ㎓ EFT/B: 2 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| EN IEC 61000-6-3:2021 |
Electromagnetic compatibility (EMC) - Part 6-3: Generic standards - Emission standard for equipment in residential environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A |
소재지-4 |
| EN IEC 61000-6-4:2019 |
Electromagnetic compatibility (EMC) - Part 6-4: Generic standards - Emission standard for industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| EN 61204-3:2018 |
Low-voltage switch mode power supplies - Part 3: Electromagnetic compatibility (EMC) |
CE: 0.15 ㎒ ~ 30 ㎒ RE, RFI: 30 ㎒ ~ 1 000 ㎒ F/H: ≤16 A ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 2 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| EN 61326-3-1:2017 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-1: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - General industrial applications |
ESD: 8 ㎸ RS: 20 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 3 ㎸ Surge: 4 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-4 |
| EN 61326-3-2:2018 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-2: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - Industrial applications with specified electromagnetic environment |
ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 2 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MS: 100 A/m Dips: 30 % ~ 100 % |
소재지-4 |
| EN 61547:2009 |
Equipment for general light purposes - EMC immunity requirements |
ESD: 4 ㎸ ~ 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 1 ㎸ Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: (0 ~ 100 ) % |
소재지-4 |
| EN 62040-2:2018 |
Uninterruptible power systems (UPS) - Part 2: Electromagnetic compatibility (EMC) requirements |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 1 ㎸ Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m |
소재지-4 |
| EN IEC 61000-3-2:2019 |
ElectroMagnetic Compatibility(EMC) - Part 3-2: Limitation for harmonics current emissions (equipment input current <= 16 A per phase) <제외항목> 3 phase connection equipment |
≤16 A |
소재지-4 |
| EPRI TR-102323:1997 |
Guidelines for Electromagnetic Interference Testing in Power Plants B-3.3 Surge B-3.4 Fast transient and Impulse tests B-3.5 Electrostatic discharge (ESD) |
Surge: 3 ㎸ ESD: 15 ㎸ EFT: 3 ㎸ |
소재지-4 |
| ETSI EN 300 386 V2.1.1:2016 |
Telecommunication network equipment; ElectroMagnetic Compatibility (EMC) requirements |
RE: 30 ㎒ ~ 1 ㎓ CE: 150 ㎑ ~ 30 ㎒ ESD: Contact ± 6 ㎸ Air ± 8 ㎸ RS: 10 V/m EFT: ± 1 ㎸ SURGE: ± 2 ㎸ CS: 3 V V-dip: 100 %, 30 % H/F: 16 A 이하 |
소재지-4 |
| ETSI EN 301 489-1 V2.2.3:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % CTI : Pulse 1, 2a, 2b, 3a, 3b, 4 |
소재지-4 |
| ETSI EN 301 489-13:2002 |
Electromagnetic compatibility and Radio spectrum Matters (ERM); ElectroMagnetic Compatibility(EMC) standard for radio equipment and services; Part 13: Specific conditions for Citizens Band (CB) radio and ancillary equipment (speech and non-speech) |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-15 V2.2.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 15: Specific conditions for commercially available amateur radio equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-17 V3.2.4:2020 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 17: Specific conditions for Broadband Data Transmission Systems |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-19 V2.2.1:2022 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 19: Specific conditions for Receive Only Mobile Earth Stations (ROMES) operating in the 1,5 ㎓ band providing data communications and GNSS receivers operating in the RNSS band providing positioning, navigation, and timing data; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-2 V2.1.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 2: Specific conditions for radio paging equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-3 V2.3.2:2023 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 3: Specific conditions for Short Range Devices (SRD) operating on frequencies between 9 ㎑ and 246 ㎓; Harmonised Standard for ElectroMagnetic Compatibility |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-33 V2.2.1(2019-04) |
Electromagnetic Compatibility(EMC) Standard for radio equipment and services ; Part 33:specific conditions for Ultra-Wideband(UWB) devices: Harmonized Standard covering the essential requirements of article 3.1(b) of Directive 2014/53/EU |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-34 V2.1.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 34: Specific conditions for External Power Supply (EPS) for mobile phones |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-4 V3.3.1:2021 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 4: Specific conditions for fixed radio links and ancillary equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-5 V2.2.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 5: Specific conditions for Private land Mobile Radio (PMR) and ancillary equipment (speech and non-speech) and Terrestrial Trunked Radio (TETRA) |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| ETSI EN 301 489-51 V2.1.1(2019-04) |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 51: Specific conditions for Automotive, Ground based Vehicles and Surveillance Radar Devices using 24,05 ㎓ to 24,25 ㎓, 24,05 ㎓ to 24,5 ㎓, 76 ㎓ to 77 ㎓ and 77 ㎓ to 81 ㎓; Harmonised Standard covering the essential requirements of article 3.1(b) of Directive 2014/53/EU |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % CTI : Pulse 1, 2a, 2b, 3a, 3b, 4 |
소재지-4 |
| ETSI EN 301 489-9 V2.1.1:2019 |
ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 9: Specific conditions for wireless microphones, similar Radio Frequency (RF) audio link equipment, cordless audio and in-ear monitoring devices |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A, ESD: 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1 ㎸, Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 0 % ~ 100 % |
소재지-4 |
| FCC Part 15:2020 |
Radio Frequency Devices <제외항목> 15.115 TV interfaces devices, including cable system terminal devices 15.117 TV Broadcast receivers |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 ㎓ |
소재지-4 |
| FCC Part 18:2020 |
Industrial, Scientific and Medical Equipment <제외항목> Radiated Emission test at 30 m, 300m and 1 600m distance. |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 40 ㎓ |
소재지-4 |
| IACS URE 10 Rev.7:2018 |
International association of classification societies E10 3. External power supply failure 4. Power supply variations 13. Electrostatic discharge 14. Electromagnetic field 15. Conducted Low Frequency 16. Conducted Radio Frequency 17. Burst/Fast Transients 18. Surge/voltage 19. Radiated Emission 20. Conducted Emission |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 6 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: 0% ~ 100% Conducted low frequency: 50 ㎐ ~ 12 ㎑ |
소재지-4 |
| IC RSS-210:2019/A1:2020 |
Licence-Exempt Radio Apparatus: Category I Equipment |
RE: 0.009 ㎒ ~ 40 ㎓ |
소재지-4 |
| ICES-001:2020 |
Industrial, Scientific and Medical (ISM) Equipment <제외항목> Radiated Emission test at 10 m distance. |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 1 ㎓ |
소재지-4 |
| ICES-003:2019 |
Information Technology Equipment (Including Digital Apparatus) — Limits and Methods of Measurement |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 000 ㎒ |
소재지-4 |
| IEC 60533:2015 |
Electrical and electronic installations in ships - Electromagnetic compatibility <제외항목> 7.2 Minimum immunity requirements, Table 4 - Minimum immunity requirements for ship equipment, AC and DC power Port, Conducted low frequency interference |
RE: 150 ㎑ ~ 2 ㎓ CE: 10 ㎑ ~ 30 ㎒ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 20 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-4 |
| IEC 60571:2012 |
Railway applications – Electronic equipment used on rolling stock 5.1 Power supply 5.2 and 12.2.6 Supply over voltages 5.4 and 12.2.7 Surge, electrostatic discharge (ESD) and transient burst susceptibility tests 5.5 Electromagnetic compatibility tests 12.2.8 Radio interference test |
O.V: AC/DC 300 V SURGE: 7.4 ㎸ ESD: 8 ㎸, BURST: 4.4 ㎸ RS: 20 V/m, 80 ㎒ ~ 2.5 ㎓ RE: 30 ㎒ ~ 1 ㎓ CE: 9 ㎑ ~ 30 ㎒ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 40 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-4 |
| IEC 60601-1-2:2014 |
Medical electrical equipment - General requirement for safety - Collateral standard : Electromagnetic compatibility Requirements and test |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-4 |
| IEC 60601-1-2:2014/A1:2020 |
General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 15 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ 27 V/m, 385 ㎒ 28 V/m, 450 ㎒, 810 ㎒, 870 ㎒, 930 ㎒, 1 720 ㎒, 1 845 ㎒, 1 970 ㎒, 2 450 ㎒ 9V/m, 710 ㎒, 740 ㎒, 780 ㎒, 5 240 ㎒, 5 500 ㎒, 5 785 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 6V, 0.15 ㎒ ~ 80 ㎒ MS: 30 A/m Dips: 30 % ~ 100 % |
소재지-4 |
| IEC 60945:2002 |
Maritime navigation and radio communication equipment and systems General requirements 4.3. Power supply 7. Power supply-Methods of testing and required test results 9. Electromagnetic emission 10. Immunity to electromagnetic environment |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸ SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 20 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-4 |
| IEC 61000-3-2:2018 |
ElectroMagnetic Compatibility(EMC) - Part 3-2: Limits for harmonic current emissions (equipment input current <= 16 A per phase) <제외항목> 3 phase connection equipment |
≤16 A |
소재지-4 |
| IEC 61000-3-2:2018/A1:2020 |
Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic current emissions (equipment input current ≤16 A per phase) |
≤16 A |
소재지-4 |
| IEC 61000-3-3:2013/A1:2017 |
ElectroMagnetic Compatibility(EMC) - Part 3-3: Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems, for equipment with rated current <= 16 A per phase and not subject to conditional connection <제외항목> 3 phase connection equipment |
≤16 A |
소재지-4 |
| IEC 61000-4-11:2004/A1:2017 |
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests |
0 % ~ 100 % |
소재지-4 |
| IEC 61000-4-11:2020 |
Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests for equipment with input current up to 16 A per phase |
0 % ~ 100 % |
소재지-4 |
| IEC 61000-4-2:2008 |
Electromagnetic Compatibility(EMC) - Part 4-2: Testing and measurement techniques - Electromagnetic discharge immunity test |
30 ㎸ |
소재지-4 |
| IEC 61000-4-3:2006/A1:2007/A2:2010 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| IEC 61000-4-3:2020 |
Electromagnetic compatibility (EMC) – Part 4-3: Testing and measurement techniques – Radiated, radio-frequency, electromagnetic field immunity test |
20 V/m, 80 ㎒ ~ 1 000 ㎒ 10 V/m, 1 000 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| IEC 61000-4-39:2017 |
Electromagnetic compatibility (EMC) – Part 4-39: Testing and measurement techniques - Radiated fields in close proximity - Immunity test |
9 ㎑ ~ 300 ㎒ / Only used section 6.1 for magnetic field immunity |
소재지-4 |
| IEC 61000-4-4:2012 |
ElectroMagnetic Compatibility(EMC) - Part 4-4: Testing and measurement techniques - Electrical fast transient/burst immunity test |
4.0 ㎸ |
소재지-4 |
| IEC 61000-4-5:2014/A1:2017 |
Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement techniques - Surge immunity test |
4.0 ㎸ |
소재지-4 |
| IEC 61000-4-6:2013 |
ElectroMagnetic Compatibility(EMC) - Part 4-6: Testing and measurement techniques - Immunity to conducted disturbances, induced by radio - frequency fields |
0.15 ㎒ ~ 230 ㎒, 10 V |
소재지-4 |
| IEC 61000-4-8:2009 |
ElectroMagnetic Compatibility(EMC) - Part 4-8: Testing and measurement techniques - Power frequency magnetic field immunity test |
1 000 A/m |
소재지-4 |
| IEC 61000-4-9:2016 |
Electromagnetic compatibility (EMC) - Part 4-9: Testing and measurement techniques - Impulse magnetic field immunity test |
100 A/m |
소재지-4 |
| IEC 61000-6-1:2016 |
Electromagnetic compatibility (EMC) - Part 6-1: Generic standards - Immunity standard for residential, commercial and light-industrial environments |
ESD: 8 ㎸, MS: 3 A/m RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒, 1.4 ㎓ ~ 6 ㎓ EFT/B: 1 ㎸,,Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| IEC 61000-6-2:2016 |
Electromagnetic compatibility (EMC) - Part 6-2: Generic standards - Immunity standard for industrial environments |
ESD: 8 ㎸, MS: 30 A/m RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30 % ~ 100 % |
소재지-4 |
| IEC 61000-6-3:2020 |
Electromagnetic compatibility (EMC) - Part 6-3: Generic standards - Emission standard for equipment in residential environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ F/H: ≤16 A |
소재지-4 |
| IEC 61000-6-4:2018 |
Electromagnetic compatibility (EMC) - Part 6-4: Generic standards - Emission standard for industrial environments |
CE: 0.15 ㎒ ~ 30 ㎒ RE : 30 ㎒ ~ 6 000 ㎒ |
소재지-4 |
| IEC 61204-3:2016 |
Low voltage power supplies, d.c. output - Part 3: ElectroMagnetic Compatibility(EMC) |
CE: 0.15 ㎒ ~ 30 ㎒ RE, RFI: 30 ㎒ ~ 1 000 ㎒. F/H: ≤16 A, ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 700 ㎒ EFT/B: 4 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Dips: 20 % ~ 100 % |
소재지-4 |
| IEC 61326-1:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A ESD: 8 kV RS : 10 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 ㎸, Surge : 2 ㎸, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-2-1:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-1: Particular requirements - Test configurations, operational conditions and performance criteria for sensitive test and measurement equipment for EMC unprotected applications |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-2-2:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-2: Particular requirements - Test configurations, operational conditions and performance criteria for portable testing, measuring and monitoring equipment used in low-voltage distribution systems |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-2-3:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-3: Particular requirements - Test configuration, operational conditions and performance criteria for transducers with integrated or remote signal conditioning <제외항목> Radiated Emission test at 10 m distance. |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-2-4:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-4: Particular requirements - Test configurations, operational conditions and performance criteria for insulation monitoring devices according to IEC 61557-8 and for equipment for insulation fault location according to IEC 61557-9 |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, Burst : 2 kV, Surge : 2 kV, CS : 10 V, 0.15 ㎒ ~ 80 ㎒ MF : 30 A/m, Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-2-5:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-5: Particular requirements - Test configurations, operational conditions and performance criteria for field devices with field bus interfaces according to IEC 61784-1 |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A, ESD: 8 kV RS : 10 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 2 kV, Surge : 2 kV, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-2-6:2020 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment |
CE: 0.009 ㎒ ~ 30 ㎒ RE: 0.009 ㎒ ~ 18 ㎓ F/H: ≤16 A ESD: 15 ㎸ RS : 10 V/m, 80 ㎒ ~ 6 ㎓, Burst : 2 kV, Surge : 2 kV, CS : 6 V, 0.15 ㎒ ~ 80 ㎒ MF : 30 A/m, Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-3-1:2017 |
Electrical equipment for measurement, control and laboratory use - EMC requirements – Part 3-1: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) – General industrial applications |
ESD: 8 kV RS : 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 30 A/m, Burst : 3 kV, Surge : 4 kV, CS : 10 V, Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61326-3-2:2017 |
Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 3-2: Immunity requirements for safety-related systems and for equipment intended to perform safety-related functions (functional safety) - Industrial applications with specified electromagnetic environment |
ESD: 8 kV RS : 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 6 ㎓, MF : 100 A/m, Burst : 2 kV, Surge : 2 kV, CS : 10 V, Dip : 30 ~ 100 % |
소재지-4 |
| IEC 61547:2009 |
Equipment for general lighting purposes - EMC immunity requirements |
ESD: 4 ㎸ ~ 8 ㎸ RS: 3 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 1 ㎸ Surge: 2 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: (0 ~ 100) % |
소재지-4 |
| IEC 61547:2020 |
Equipment for general lighting purposes - EMC immunity requirements |
ESD: 8 ㎸ RS : 80 ㎒ ~ 1 ㎓, MF : 3 A/m, Burst : 1 ㎸, Surge : 2 ㎸, CS : 3 V, 0.15 ㎒ ~ 80 ㎒ Dip : 0 % ~ 100 % |
소재지-4 |
| IEC 62040-2:2016 |
Uninterruptible power system (UPS) Part 2: ElectroMagnetic Compatibility(EMC) requirements <제외항목> 7.4 Immunity to low-frequency signals |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, MS: 30 A/m RS: 10 V/m, 80 ㎒ ~ 1 000 ㎒ EFT/B: 2 ㎸, Surge: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ |
소재지-4 |
| IEC 62236-2:2018 |
Railway applications - Electromagnetic compatibility - Part 2: Emission of the whole railway system to the outside world <제외항목> Radiated Emission test at 10 m distance. |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-4 |
| IEC 62236-3-1:2018 |
Railway applications - Electromagnetic compatibility - Part 3-1: Rolling stock - Train and complete vehicle <제외항목> Radiated Emission test at 10 m distance. |
RE: 9 ㎑ ~ 1 ㎓ |
소재지-4 |
| IEC 62236-3-2:2018 |
Railway applications - Electromagnetic compatibility - Part 3-2: Rolling stock - Apparatus |
CE: 0.15 ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Burst: 2 ㎸, Surge: 2 ㎸, RS: 20 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| IEC 62236-4:2018 |
Railway applications - Electromagnetic compatibility Part 4: Emission and immunity of the signalling and telecommunications apparatus |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, SURGE : ± 2 ㎸ RS: 20 V/m, 80 ㎒ ~ 6 ㎓ EFT: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m |
소재지-4 |
| KRS CS 0003-13 (R) |
철도차량 차상신호장치 시험방법 4.3절 아. 전기자기 적합성 시험 |
CE: 0.15 ~ 30 ㎒ RE: 30 ㎒ ~ 1 000 ㎒ ESD: 8 ㎸, CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Burst: 2 ㎸, Surge: 2 ㎸, RS: 20 V/m 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KRS SG 0015-20 (R) |
전자연동장치 4.7.3 전기자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KRS SG 0033-21 (R) |
유절연가청주파수(AF) 궤도회로 4.2.2절 6) 전기자기적합성 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KRS SG 0038-21 (R) |
무절연가청주파수(AF) 궤도회로 4.5.3 전기자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KRS SG 0054-21 (R) |
단선 자동폐색제어장치 4.2.2절 6) 전자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KRS SG 0055-21 (R) |
복선 자동폐색제어장치 4.2.2절 6) 전자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KRS SG 0067-14 (R) |
지상 신호제어설비 시험방법 4.3.4 전기자기 적합성(EMC) 시험 |
CE: 0.15 ㎑ ~ 30 ㎒ RE: 30 ㎒ ~ 6 ㎓ ESD: 8 ㎸, Surge : ± 2 ㎸ Burst: 2 ㎸ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ MF: 0, 16.7, 50, 60 ㎐ / 300 A/m RS: 20 V/m, 80 ㎒ ~ 1 ㎓, 1.4 ㎓ ~ 2.7 ㎓, 5.1 ㎓ ~ 6 ㎓ |
소재지-4 |
| KR:2020 |
제조법 및 형식승인 등에 관한 지침 23. 자동화시스템 3. 전원상실시험 4. 전원변동시험 14. 정전기방전내성시험 15. 방사전자계 내성시험 16. 전도성 저주파 간섭에 대한 내성시험 17. 전도성 고주파 간섭에 대한 내성시험 18. 버스트 내성시험 19. 서지 내성시험 20. 발산방사 측정시험 21. 전도방사 측정시험 |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 6 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: 0% ~ 100% Conducted low frequency: 50 ㎐ ~ 12 ㎑, 1~10% or Max 2 W |
소재지-4 |
| KS C IEC 60571:2012 |
철도 적용 - 철도 차량용 전자기기 5.1 공급 전원 5.2 와 12.2.6 과전압 5.4 와 12.2.7 서지, 정전기와 전기적 빠른 과도현상/버스트 내성시험 5.5 전자기 적합성 12.2.8 무선 장해 |
O.V: AC/DC 300 V SURGE: 7.4 ㎸ ESD: 8 ㎸ RS: 20 V/m 80 ㎒ ~ 2.5 ㎓ BURST: 4.4 ㎸ RE: 30 ㎒ ~ 1 ㎓ CE: 9 ㎑ ~ 30 ㎒ CS: 10 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 40 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-4 |
| KS X IEC 60945:2002 |
해상 항해, 무선통신 기기 및 시스템 - 일반 요구사항 - 시험 방법과 요구되는 시험결과 4.3. 공급 전원 7. 공급 전원 시헙방법 및 요구되는 결과 9. 전자파 장해 10. 전자파내성 |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸ SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: V 20 % Variation (1.5 s) F 10 % Variation (5 s) V 100 % Dip (60 s) |
소재지-4 |
| LR Type approval system:2020 |
LR Type approval system - Test specification number1 2013 - Electrical equipment control and monitoring equipment instrumentation and internal communication equipment computer and peripheral equipment 8. Power supply variation test 9. Power supply failure test 22. Immunity to conducted high frequency interference 23. Immunity to conducted low frequency interference 24. Immunity to radiated radio frequency field 25. Immunity to fast low energy transients (bursts) 26. Immunity to slow high energy transients (surges) 27. Immunity to electrostatic discharge (ESD) 29. Radiated emissions 30. Conducted emissions |
CE: 10 ㎑ ~ 30 ㎒ RE: 150 ㎑ ~ 2 ㎓ ESD: 8 ㎸ RS: 10 V/m, 80 ㎒ ~ 2 000 ㎒ EFT: 2 ㎸, SURGE: 1 ㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Voltage Variation: Voltage 20 % Variation (1.5 s) Freq. 10 % Variation (5 s) Voltage 100 % Dip (60 s) Conducted low frequency: 50 ㎐ ~ 12 ㎑, 1 % ~ 10 % or Max 2 W |
소재지-4 |
| MIL-STD-461D:1993 |
REQUIREMENTS FOR THE CONTROL OF ELECTROMAGNETIC INTERFERENCE EMISSIONS AND SUSCOPTIBILITY 5.3.1 CE101, Conducted emissions, power leads, 30 ㎐ to 10 ㎑ 5.3.2 CE102, Conducted emissions, power leads, 10 ㎑ to 10 ㎒ 5.3.3 CE106, Conducted emissions, antenna terminal, 10 ㎑ to 40 ㎓ 5.3.4 CS101, Conducted susceptibility, power leads, 30 ㎐ to 50 ㎑ 5.3.5 CS103, Conducted susceptibility, antenna port, intermodulation, 15 ㎑ to 10 ㎓ 5.3.6 CS104, Conducted susceptibility, antenna port, rejection of undesired signals, 30 ㎐ to 20 ㎓ 5.3.7 CS105, Conducted susceptibility, antenna port, cross modulation, 30 ㎐ to 20 ㎓ 5.3.9 CS114, Conducted susceptibility, bulk cable injection, 10 ㎑ to 400 ㎒ 5.3.10 CS115, Conducted susceptibility, bulk cable injection, impulse excitation 5.3.11 CS116, Conducted susceptibility, damped sinusoidal transients, cables and power leads, 10 ㎑ to 100 ㎒ 5.3.12 RE101, Radiated emissions, magnetic field, 30 ㎐ to 100 ㎑ 5.3.13 RE102, Radiated emissions, electric field, 10 ㎑ to 18 ㎓ 5.3.15 RS101, Radiated susceptibility, magnetic field, 30 ㎐ to 100 ㎑ 5.3.16 RS103, Radiated susceptibility, electric field, 10 ㎑ to 18 ㎓ <제외항목> 10 ㎑ to 80 ㎒, 200 V/m, 1 m distance |
CE101: 30 ㎐ ∼ 10 ㎑ CE102: 10 ㎑ ∼ 10 ㎒ CE106: 10 ㎑ ~ 40 ㎓ CS101: 30 ㎐ ∼ 50 ㎑ CS103: 15 ㎑ ~ 10 ㎓ CS104: 30 ㎐ ~ 20 ㎓ CS105: 30 ㎐ ~ 20 ㎓ CS114: 10 ㎑ ∼ 400 ㎒ CS115: Impulse 5 A CS116: 10 ㎑ ∼ 100 ㎒ RE101: 30 ㎐ ∼ 100 ㎑ RE102: 10 ㎑ ∼ 18 ㎓ RS101: 30 ㎐ ∼ 100 ㎑ RS103: 10 ㎑ ∼ 18 ㎓ (Max. 200 V/m) |
소재지-4 |
| MIL-STD-461E:1999 |
REQUIREMENTS FOR THE CONTROL OF ELECTROMAGNETIC INTERFERENCE CHARACTERISTICS OF SUBSYSTEMS AND EQUIPMENT 5.4 CE101, conducted emissions, power leads, 30 ㎐ to 10 ㎑ 5.5 CE102, conducted emissions, power leads, 10 ㎑ to 10 ㎒ 5.6 CE106, conducted emissions, antenna terminal, 10 ㎑ to 40 ㎓ 5.7 CS101, conducted susceptibility, power leads, 30 ㎐ to 150 ㎑ 5.8 CS103, conducted susceptibility, antenna port, intermodulation, 15 ㎑ to 10 ㎓ 5.9 CS104, conducted susceptibility, antenna port, rejection of undesired signals, 30 ㎐ to 20 ㎓ 5.10 CS105, conducted susceptibility, antenna port, cross modulation, 30 ㎐ to 20 ㎓ 5.12 CS114, conducted susceptibility, bulk cable injection, 10 ㎑ to 200 ㎒ 5.13 CS115, conducted susceptibility, bulk cable injection, impulse excitation 5.14 CS116, conducted susceptibility damped sinusoidal transients, cable and power leads, 10 ㎑ to 100 ㎒ 5.15 RE101, radiated emissions, magnetic field 30 ㎐ to 100 ㎑ 5.16 RE102, radiated emissions, electric field, 10 ㎑ to 18 ㎓ 5.18 RS101, radiated susceptibility, magnetic field, 30 ㎐ to 100 ㎑ 5.19 RS103, radiated susceptibility, electric field, 2 ㎒ to 18 ㎓ <제외항목> 2 ㎒ to 80 ㎒, 200 V/m, 1 m distance |
CE101: 30 ㎐ ∼ 10 ㎑ CE102: 10 ㎑ ∼ 10 ㎒ CE106: 10 ㎑ ~ 40 ㎓ CS101: 30 ㎐ ∼ 50 ㎑ CS103: 15 ㎑ ~ 10 ㎓ CS104: 30 ㎐ ~ 20 ㎓ CS105: 30 ㎐ ~ 20 ㎓ CS114: 10 ㎑ ∼ 400 ㎒ CS115: Impulse 5 A CS116: 10 ㎑ ∼ 100 ㎒ RE101: 30 ㎐ ∼ 100 ㎑ RE102: 10 ㎑ ∼ 18 ㎓ RS101: 30 ㎐ ∼ 100 ㎑ RS103: 10 ㎑ ∼ 18 ㎓ (Max. 200 V/m) |
소재지-4 |
| MIL-STD-461F:2007 |
REQUIREMENTS FOR THE CONTROL OF ELECTROMAGNETIC INTERFERENCE CHARACTERISTICS OF SUBSYSTEMS AND EQUIPMENT 5.4 CE101, conducted emissions, power leads, 30 ㎐ to 10 ㎑ 5.5 CE102, conducted emissions, power leads, 10 ㎑ to 10 ㎒ 5.6 CE106, conducted emissions, antenna terminal, 10 ㎑ to 40 ㎓ 5.7 CS101, conducted susceptibility, power leads, 30 ㎐ to 150 ㎑ 5.8 CS103, conducted susceptibility, antenna port, intermodulation, 15 ㎑ to 10 ㎓ 5.9 CS104, conducted susceptibility, antenna port, rejection of undesired signals, 30 ㎐ to 20 ㎓ 5.10 CS105, conducted susceptibility, antenna port, cross modulation, 30 ㎐ to 20 ㎓ 5.11 CS106, conducted susceptibility, transients, power leads 5.13 CS114, conducted susceptibility, bulk cable injection, 10 ㎑ to 200 ㎒ 5.14 CS115, conducted susceptibility, bulk cable injection, impulse excitation 5.15 CS116, conducted susceptibility, damped sinusoidal transients, cable and power leads, 10 ㎑ to 100 ㎒ 5.16 RE101, radiated emissions, magnetic field, 30 ㎐ to 100 ㎑ 5.17 RE102, radiated emissions, electric field, 10 ㎑ to 18 ㎓ 5.19 RS101, radiated susceptibility, magnetic field, 30 ㎐ to 100 ㎑ 5.20 RS103, radiated susceptibility, electric field, 2 ㎒ to 18 ㎓ <제외항목> 2 ㎒ to 80 ㎒, 200 V/m, 1 m distance |
CE101: 30 ㎐ ∼ 10 ㎑ CE102: 10 ㎑ ∼ 10 ㎒ CE106: 10 ㎑ ∼ 40 ㎓ CS101: 30 ㎐ ∼ 150 ㎑ CS103: 15 ㎑ ~ 10 ㎓ CS104: 30 ㎐ ~ 20 ㎓ CS105: 30 ㎐ ~ 20 ㎓ CS106: 400 Vpeak CS114: 4 ㎑ ∼ 200 ㎒ CS115: Impulse 5 A CS116: 10 ㎑ ∼ 100 ㎒ RE101: 30 ㎐ ∼ 100 ㎑ RE102: 10 ㎑ ∼ 18 ㎓ RS101: 30 ㎐ ∼ 100 ㎑ RS103: 2 ㎒ ∼ 18 ㎓ (Max. 200 V/m) |
소재지-4 |
| MIL-STD-461G:2015 |
REQUIREMENTS FOR THE CONTROL OF ELECTROMAGNETIC INTERFERENCE CHARACTERISTICS OF SUBSYSTEMS AND EQUIPMENT 5.4 CE101, conducted emissions, audio frequency currents, power leads 5.5 CE102, conducted emissions, radio frequency potential, power leads 5.6 CE106, conducted emissions, antenna port 5.7 CS101, conducted susceptibility, power leads 5.8 CS103, conducted susceptibility, antenna port, intermodulation 5.9 CS104, conducted susceptibility, antenna port, rejection of undesired signals 5.10 CS105, conducted susceptibility, antenna port, cross modulation 5.12 CS114, conducted susceptibility, bulk cable injection 5.13 CS115, conducted susceptibility, bulk cable injection, impulse excitation 5.14 CS116, conducted susceptibility, damped sinusoidal transients, cables and power leads 5.16 CS118, personnel borne electrostatic discharge 5.17 RE101, radiated emissions, magnetic field 5.18 RE102, radiated emissions, electric field 5.20 RS101, radiated susceptibility, magnetic field 5.21 RS103, radiated susceptibility, electric field <제외항목> 2 ㎒ to 80 ㎒, 200 V/m, 1 m distance |
CE101: 30 ㎐ ∼ 10 ㎑ CE102: 10 ㎑ ∼ 10 ㎒ CE106: 10 ㎑ ~ 40 ㎓ CS101: 30 ㎐ ∼ 150 ㎑ CS103: 15 ㎑ ~ 10 ㎓ CS104: 30 ㎐ ~ 20 ㎓ CS105: 30 ㎐ ~ 20 ㎓ CS114: 4 ㎑ ∼ 200 ㎒ CS115: Impulse 5 A CS116: 10 ㎑ ∼ 100 ㎒ CS118: Contact/Air, (±2 ~ ±15) kV RE101: 30 ㎐ ∼ 100 ㎑ RE102: 10 ㎑ ∼ 18 ㎓ RS101: 30 ㎐ ∼ 100 ㎑ RS103: 2 ㎒ ∼ 18 ㎓ (Max. 200 V/m) |
소재지-4 |
| RSS-Gen:2014 |
General Requirements and Information for the Certification of Radio Apparatus |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 ㎓ |
소재지-4 |
| RSS-Gen:2018/A1:2019/A2:2021 |
General Requirements for Compliance of Radio Apparatus |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 40 ㎓ |
소재지-4 |
| TEC/SD/DD/EMC-221/05/OCT-16 |
Electromagnetic compatibility Standard for Telecommunication Equipment |
CE: 0.15 ㎒ ~ 30 ㎒ RE: 30 ㎒ ~ 6 000 ㎒ ESD: 8㎸ RS: 10 V/m, 80 ㎒ ~ 6 000 ㎒ EFT/B: 1㎸, Surge: 2㎸ CS: 3 V, 0.15 ㎒ ~ 80 ㎒ Dips: 30%~ >95% |
소재지-4 |
| U.SNRC RG-1.10:2003 |
Guidelines for evaluating electromagnetic and radio-frequency interference in safety-related instrumentation and control system - Application Below 100 A <제외항목> Table 7 - IEC 61000-4-10, 12, 13, 16 |
CE: 30 ㎐ ~ 10 ㎒ RE: 30 ㎐ ~ 18 ㎓ CS: 30 ㎐ ~ 200 ㎒ RS: 30 ㎐ ~ 18 ㎓ SURGE: 6 ㎸ EFT: 4 ㎸, ESD: 15 ㎸ |
소재지-4 |
| U.SNRC RG-1.180:2003 |
Guidelines for evaluating electromagnetic and radio-frequency interference in safety-related instrumentation and control system - Application Below 100 A <제외항목> Table 7 - IEC 61000-4-10, 12, 13, 16 |
CE: 30 ㎐ ~ 10 ㎒ RE: 30 ㎐ ~ 18 ㎓ CS: 30 ㎐ ~ 200 ㎒ RS: 30 ㎐ ~ 18 ㎓ SURGE: 6 ㎸ EFT: 4 ㎸, ESD: 15 ㎸ |